Performance Check— Type 1A1
d. Connect a patch cord between Channel 1 INPUT and
the Time Base B + Gate Out connector.
e. Turn the test oscilloscope Intensity control counter
clockwise to reduce the intensity to normal brilliance to
avoid burning the CRT. The display obtained should be
a falling RC curve as shown in Fig.
6
-
8
.
f. Adjust the Channel 1 VARIABLE VOLTS/CM control
for a display amplitude of 4 cm.
g. Position the display to the center of the graticule area
and measure the time it takes the waveform to fall from
4 cm to 1.5 cm (see Fig.
6
-
8
). It should be more time than
0.008 sec (less than
2
hertz).
h. Change the patch cord to Channel 2 INPUT and set
the MODE switch to CH 2.
i. Repeat the procedure to check the Channel 2 low-
frequency response.
j. Remove the patch cord. This completes the perform
ance checkout procedure for the Type 1A1 Dual-Trace Plug-
In Unit. If the instrument has met all performance require
ments given in this procedure, it is correctly calibrated and
within the specified tolerances.
6-11
Содержание 1A1
Страница 47: ...Maintenance Type 1A1 670 0075 01 Fig 5 5 Channel 1 Input Amplifier card 5 13 http manoman sqhill com ...
Страница 50: ...Maintenance Type 1AT Fig 5 8 Front of Output Amplifier card 5 16 http manoman sqhill com ...
Страница 128: ...ALT SWEEP SLAVE PULSE CHOPPED BLANKING ALT TRACE SYNC PULSE N 2 0 0 0 0 U P SWITCHING CIRCUIT TYPE IAI PLUG IN ...
Страница 129: ...http manoman sqhill com ...
Страница 130: ...38 ...
Страница 131: ...FIG 3 CIRCUIT CARDS ACCESSORIES CHANNEL I INPUT OUTPUT AMPLIFIER CHANNEL 2 INPUT TYPE 1A1 DUAL TRACE PLUG IN UNIT ...
Страница 134: ...Page 2 o f 2 TYPE 1A1 A p a r t i a l CHAKJKJEL Z INPUT A M P L I F I E R M16 025 1269 ...