Performance Check— Type 1A1
1
/
R
i
k K
4 cm
Optimum
and Fiat
Squc
Top
>eak-<
ire Co
within
o-Peo
rner
<
71 1.6 mm
r
-
-
Sweep Rate:
0.1 /zsec/cm
(A ) Channel 1.
L
i
l
l
i
- Optimum Square Corner >
R
1.6 mm Peak-to -Peak
k |
i
-
-
i
-
-
Sweep Rate:
0.1 /zsec/cm
(B) Channel 2.
Fig. 6-5. Typical Channel 1 and Channel 2 waveforms of high-
frequency response at a vertical deflection factor of 0.0 05 volts/cm.
should not exceed 1.6 mm (or 4%) peak to peak for a 4-cm
amplitude display.
h.
Remove the 10X attenuator. Apply the 100 kHz fast-
rise signal from the +O utp ut connector on the Type 106
through the 5-ns cable and 50-ohm in-line termination to
the Type 1A1 Channel 1 INPUT connector.
17. Check Trigger Out Amplifier Risetime
a. Requirement— Risetime of the Trigger Out Amplifier
waveform not to exceed 70-ns.
b. Set Channel 1 VOLTS/CM switch to .05, Channel 2
VOLTS/CM switch to .5, and the Channel 2 INPUT SELECTOR
switch to AC.
c. Connect a 50-ohm coaxial cable from the CH 1 TRIG
GER OUT connector to the CH 2 INPUT connector.
d. Check that the MODE switch is set to CH 2. Set the
Type 106 -(-Transition Amplitude control to obtain a 4-cm
amplitude display. Use the Channel 2 POSITION control
to center the display. Set the oscilloscope Sweep Magnifier
switch to X5.
e. CHECK— Risetime of the trigger out waveform must
be equal to or less than 70-ns (see Fig.
6
-
6
).
f. Disconnect the Type 106 signal from the Channel 1 and
turn off the generator. Remove the 50-ohm coaxial cable
from the CH 1 TRIGGER OUT and Channel 2 INPUT con
nectors.
Fig. 6 -6 . Measuring the ristime of the CH 1 TRIGGER OUT wave
form. Risetime should be 70 ns or less; waveform shown above has
a risetime of 40 ns.. Sweep rate: 20 nsec/cm.
18. Check Upper Bandwidth Limit (Both
Channels)
a. Requirement— 50 MHz or greater at the 3 dB down
point. (See Table 6-3 if oscilloscope is not a Type 547.)
b. Set the equipment controls as follows:
Type 1A1
MODE
CH 1
POSITION
(both channels)
Near midrange
PULL FOR INVERT
(both channels)
Normal or in
VARIABLE VOLTS/CM
(both channels)
CALIB
VOLTS/CM
(both channels)
.05
INPUT SELECTOR
(channel
2
)
DC
Type
B Time/cm
Sweep Magnifier
Triggering Level
Horizontal Position
547
.5 mSec
X I O ff
Fully clockwise
Trace positioned to start
at left of graticule line
Type 191 Constant Amplitude Signal Generator
Frequency dial
Frequency Range
Amplitude
Variable
Amplitude Range
Power
45 MHz
50 kHz Only
20
Cal
50-500 mV
On
c.
Apply the 50-kHz reference signal from the Type 191
Output connector through a 5-ns cable and a 50-ohm in-line
termination to the Type 1A1 Channel 1 INPUT connector.
6-8
@2
Содержание 1A1
Страница 47: ...Maintenance Type 1A1 670 0075 01 Fig 5 5 Channel 1 Input Amplifier card 5 13 http manoman sqhill com ...
Страница 50: ...Maintenance Type 1AT Fig 5 8 Front of Output Amplifier card 5 16 http manoman sqhill com ...
Страница 128: ...ALT SWEEP SLAVE PULSE CHOPPED BLANKING ALT TRACE SYNC PULSE N 2 0 0 0 0 U P SWITCHING CIRCUIT TYPE IAI PLUG IN ...
Страница 129: ...http manoman sqhill com ...
Страница 130: ...38 ...
Страница 131: ...FIG 3 CIRCUIT CARDS ACCESSORIES CHANNEL I INPUT OUTPUT AMPLIFIER CHANNEL 2 INPUT TYPE 1A1 DUAL TRACE PLUG IN UNIT ...
Страница 134: ...Page 2 o f 2 TYPE 1A1 A p a r t i a l CHAKJKJEL Z INPUT A M P L I F I E R M16 025 1269 ...