Calibration— Type 1A1
f. CHECK— For proper vertical deflection at each Channel
1 VOLTS/CM switch position by using Table 7-1 as a con-
convenient guide. Accuracy is ± 3 % for the .01, .02 and .1
through
20
switch positions when the gain is accurately set at
the .005 and .05 positions.
g. After checking the attenuation ratios of the Channel 1
VOLTS/CM switch, set the Standard Amplitude Calibrator
for a 20 mV output.
h. Apply the calibrator signal to the Channel 2 INPUT
connector.
i. Set the MODE switch to CH 2.
j. CHECK— For proper vertical deflection at each Chan
nel 2 VOLT/CM switch position. Use Table 7-1 as a guide.
k. After completing the previous step, turn off the Stand
ard Amplitude Calibrator and disconnect the 50-ohm coaxial
cable.
TABLE 7-1
VOLTS/CM
Switch
Setting
Standard
Am plitude
C alibrator O utput
(peak to peak)
Vertical
Vertical
Deflec
tion in
cm
Maximum
Error for
± 3 %
Accuracy
.005
20 mVolts
4
Adjusted
0
.01
50 mVolts
5
± 1 .5 mm
.02
.1 Volts
5
± 1 .5 mm
.05
.2 Volt
4
Adjusted
0
.1
.5 Volt
5
± 1 .5 mm
.2
1 Volt
5
± 1 .5 mm
.5
2 Volts
4
±
1.2
mm
1
5 Volts
5
± 1 .5 mm
2
10 Volts
5
± 1 .5 mm
5
20 Volts
4
±
1 .2
mm
10
50 Volts
5
± 1 .5 mm
20
100 Volts
5
± 1 .5 mm
°The am plifier gain for these switch
positions has been previously
adjusted for exact amplitude. See steps 5, 7, 9 and 10.
7-21
Содержание 1A1
Страница 47: ...Maintenance Type 1A1 670 0075 01 Fig 5 5 Channel 1 Input Amplifier card 5 13 http manoman sqhill com ...
Страница 50: ...Maintenance Type 1AT Fig 5 8 Front of Output Amplifier card 5 16 http manoman sqhill com ...
Страница 128: ...ALT SWEEP SLAVE PULSE CHOPPED BLANKING ALT TRACE SYNC PULSE N 2 0 0 0 0 U P SWITCHING CIRCUIT TYPE IAI PLUG IN ...
Страница 129: ...http manoman sqhill com ...
Страница 130: ...38 ...
Страница 131: ...FIG 3 CIRCUIT CARDS ACCESSORIES CHANNEL I INPUT OUTPUT AMPLIFIER CHANNEL 2 INPUT TYPE 1A1 DUAL TRACE PLUG IN UNIT ...
Страница 134: ...Page 2 o f 2 TYPE 1A1 A p a r t i a l CHAKJKJEL Z INPUT A M P L I F I E R M16 025 1269 ...