Calibration— Type 1A1
e.
CHECK—Sot the Channel 1 INPUT SELECTOR switch to
AC and nolo the amount of trace shift; should not exceed
2 mm,
f
Disconnect the 50-ohm termination
13. Adjust Channel 1 INV BAL (R152)
O
0. Equipment setup is shown in Fig 7-13. (The 50-ohm
termination is not used )
b.
Set Channel 1 INPUT SELECTOR switch to GND and
check that the VOLTS/CM switch is set to .005.
C, Chock that the trace is positioned to the horizontal
centerline. Note the trace position
d. Set the Channel 1 PULL FOR INVERT switch to invert
e. CHECK— That there is no trace shift when switching
to the invert position, If there is, leave the switch in the
invert position
f. ADJUST—Channel I INV BAL control (see Fig 7-14) to
vertically position the froce to the point noted in step c,
NOTE
To check on the accuracy of your adjustment, set
Channel I PULL FOR INVERT switch to normal.
Then repeat steps d and e and readjust, if neces
sary, the Channel 1 INV BAL control for no trace
shift.
h. Set Channel 2 PULL FOR INVERT switch to normal.
1. INTERACTION—Dependent on proper adjustment per
formed in steps 1 ond 2.
14. Adjust Channel 2 INV BAL (R252)
O
a.
Equipment setup is shown in Fig, 7-13, (The 50-ohm
termination is not used.)
b. Set the Channel 2 INPUT SELECTOR switch to GND
ond the MODE switch to CH 2, Check that the Channel 2
VOLTS/CM switch is set to .005.
c. Check that the trace is positioned to the horizontal
centorhne Note the position of (he trace,
d. Set the Channel 2 PULL FOR INVERT switch to invert.
e. CHECK—That there is no trace shift when switching
to the invert position, If there is, leave the switch in the
invert position.
g. ADJUST— Channel 2 INV BAL control (see Fig. 7-15) to
vertically position the trace to the point noted in step c.
NOTE
To check on the accuracy of your adjustment, set
Channel 2 PULL FOR INVERT switch to normal.
Then repeat steps d and e ond readjust, if neces
sary, the Channel 2 INV BAL control for no trace
shift.
h. Set Channel 2 PULL FOR INVERT switch to normal
i. INTERACTION— Dependent on proper adiustments per
formed in steps 3 and 4.
15. Check Chopped-Mode Operation
a. Equipment sefup is shown in Fig. 7-13. (The 50-ohm
termination is not used).
b. Set the MODE switch to CHOP. Two free-running
traces should be displayed.
c. Check that both VOLTS/CM switches ore set to ,005
and the INPUT SELECTOR switches are set to GND,
d Using the Channel 1 ond 2 POSITION controls, posi
tion the Channel 1 trace above the Channel 2 trace. Separ
ate the traces so they are about two cm apart and the dis
play is centered on the CRT.
7-13
Содержание 1A1
Страница 47: ...Maintenance Type 1A1 670 0075 01 Fig 5 5 Channel 1 Input Amplifier card 5 13 http manoman sqhill com ...
Страница 50: ...Maintenance Type 1AT Fig 5 8 Front of Output Amplifier card 5 16 http manoman sqhill com ...
Страница 128: ...ALT SWEEP SLAVE PULSE CHOPPED BLANKING ALT TRACE SYNC PULSE N 2 0 0 0 0 U P SWITCHING CIRCUIT TYPE IAI PLUG IN ...
Страница 129: ...http manoman sqhill com ...
Страница 130: ...38 ...
Страница 131: ...FIG 3 CIRCUIT CARDS ACCESSORIES CHANNEL I INPUT OUTPUT AMPLIFIER CHANNEL 2 INPUT TYPE 1A1 DUAL TRACE PLUG IN UNIT ...
Страница 134: ...Page 2 o f 2 TYPE 1A1 A p a r t i a l CHAKJKJEL Z INPUT A M P L I F I E R M16 025 1269 ...