System Installation and Setup
• Set up central monitoring (if used).
• Program foreign devices into the system.
•
Save and implement the custom system configuration.
• Integrate other TekTone
®
systems such as the Tek-CARE
®
400.
•
Save the final system configuration.
• Test the system.
•
Create a backup of the system configuration.
Install the Tek-BRIDGE
™
Set up the Tek-BRIDGE™
TekTone
®
recommends that the Tek-BRIDGE™ be plugged into an uninterruptable
power supply (UPS) or dedicated life safety electrical circuit. If a UPS is required,
a TekTone
®
PK250B Uninterruptable Power Supply is recommended. Connect the
monitor, keyboard, and mouse to the Tek-BRIDGE™ and turn both the computer and
monitor on.
The Tek-BRIDGE™ is preprogrammed with a Windows
®
password. The default username
and password are as follows
:
Username:
tekbridge pc
Password:
tektone
The default system password may be changed at any time by choosing
Start>Control
Panel>User Accounts>TekBridge PC>Change my Password
or
Remove my
Password
.
After the computer finishes its startup process, verify that the system clock in the system
tray is correct. If it is not, be sure to set the correct time and date before proceeding.
Connect to the Facility’s LAN
The Tek-BRIDGE™ must be connected to the facility’s local area network (LAN) if the
system will use any of the following functions:
• Install and use
LS586 Remote Event Monitor
software on facility PCs
• View and create custom reports using facility PCs
• Use the Email Output function to send automatic emails or text messages in
response to nurse call events
Without a LAN connection, the Event Monitor and Reporting functions of the system are
only available at the Tek-BRIDGE™. The Tek-CARE
®
system includes one LS586 remote
Event Monitor license, with the ability to expand up to nine total remote Event Monitors.
For facilities that require additional licenses, contact TekTone
®
at (800) 327-8466.
LAN Connection Procedure
Connect the Tek-BRIDGE™ to the facility’s LAN using a standard network cable. The
maximum wire distance is 150’.
The Tek-BRIDGE™ is equipped with three Ethernet ports, one for Tek-CARE
®
550 system
hardware connections, a second for Tek-CARE
®
400 network connection, and a third for
the Facility LAN connection. The settings for the Facility LAN port must match those
of the host LAN. Contact the facility’s IT department for any necessary system changes
8 • IL948 Tek-CARE
®
500 & Tek-CARE
®
550 Installation Manual