Data Lo
5A
CRC
0CAB
06 (0x06) Write Single Register
11.3.5.4
Query
The Write command specifies the register reference to be preset and allows the remote
configuration of channel and system settings for management of the unit’s operation, and
modification for use in monitoring. The registers which hold the configuration parameters are:
Parameters
Registers
Decimal
Registers
Hex
Instrument Address
11
0B
Serial Type
12
0C
Here is an example of a request to preset register 11 (instrument address) of device slave 1 to
change it to devices slave 4. Note that changing the instrument address or serial type will only
take effect after the sensor is shut down and restarted:
Field Name
Example (Hex)
Slave Address
01
Function
06
Starting Address Hi
00
Starting Address Lo
0B
No. of Registers Hi
00
No. of Registers Lo
04
CRC Check
F9CB
11.3.5.5
Response
The normal response is an echo of the query, returned after the register contents have been
preset.
Here is an example of a response to above query to preset register 11, instrument address to 4.
Field Name
Example (Hex)
Slave Address
01
Function
06
Starting Address Hi
00
Starting Address Lo
0B
No. of Registers Hi
00
No. of Registers Lo
04
CRC Check
F9CB
17 (0x11) Report Slave ID
Returns information about the slave device. Currently returns only the Version No. of the probe
firmware.
Содержание OQSExISG2
Страница 1: ...UG OQSxG2 23 0025 OIL QUALITY SENSOR ASSOCIATED ACCESSORIES USER GUIDE...
Страница 2: ......
Страница 5: ...Page 3 Page Intentionally Blank...
Страница 37: ...Page 35 Figure 19 Connection to other Tan Delta and 3rd Party Products Interfaces...
Страница 38: ...Page 36 Figure 20 ExOQSxG2 Ex Sensor Installation Drawing Rev 2 STAMPED...
Страница 42: ...Page 40 8 Display Express...
Страница 97: ...Page 95 P4 CANL RS485A P5 0V P6 CANH RS485B...