Page | 121
13.1.5
Cleaning Procedure
–
Odourless Kerosene
Unplug and remove the sensor then use absorbent paper to wipe off the excess oil from
the sensor tip and thread.
Attach a bottle adaptor to a sample bottle and pour in approximately 15ml of kerosene.
Screw in the sensor and shake vigorously for 2 minutes to ensure the kerosene washes
up around the tip of the sensor.
Remove the sensor from the adaptor and shake rigorously over the absorbent paper to
dislodge any excess kerosene.
Leave to dry for a few minutes.
Figure 87 - Cleaning the Sensor (3)