DTC-300 Getting Started Guide
Page 48
4
Highlight the reference to be added. Click
Another
to continue adding reference. Click
Finished
to
continue.
Figure 23
Select Reference Material.
5
Enter three temperatures and a thickness for the number of segments that the original calibration had.
Figure 24
Setpoint information.
6
A Calibration Changes window displays, asking if you would like to create a new file to save the changes.
Click
No
to terminate the reanalysis, or click
Yes
to continue.
7
A Sample Information window displays. Update the information, if needed.
8
If the Sample/File ID field is left unchanged, a window displays asking to overwrite the file. Click
Yes
to overwrite the file or
No
to terminate modification.
9
Modified calibration results display.
Deleting a Reference
The
Delete a Reference
option allows you to delete one or more reference pieces of the calibration.
NOTE
: The calibration must always have at least 3 references.
1
In the software, click
Results
>
Post Analysis
>
Modify a Calibration File
.
2
A Calibration Modify window displays. Select
Delete a Reference
.