Choosing a Location
Because of the sensitivity of DTC-300 experiments, it is important to choose a location for the instrument
using the following guidelines. The DTC-300 system should be:
In
•
A temperature and humidity controlled area.
•
A clean, vibration-free environment.
•
An area with ample working and ventilation space.
On
•
A stable, non-flammable work surface.
Near
•
A power outlet for the specific voltage of the instrument.
•
Compressed lab air and purge gas supplies with suitable regulators and filtering where needed.
•
Cooling water/coolant (stable chiller/circulator preferred).
Away from
•
Dusty environments.
•
Exposure to direct sunlight.
•
Direct air drafts (fans, room air ducts).
•
Poorly ventilated areas.
•
Electrically noisy areas or areas prone to mechanical vibrations.
NOTE
: Do not place equipment against walls or cabinets that might impede air flow. Leave at least 7.5 cm
(3 in) clearance between the back of the instrument and any objects.
WARNING: For safety, position the equipment in a manner that allows access to
the power cord for emergency disconnection.
AVERTISSEMENT: Par mesure de sécurité, placez l'équipement de sorte qu'il per-
mette d'accéder facilement au cordon d'alimentation en cas de débranchement
d'urgence.
WARNING: Protect power and communications cable paths. Do not create
tripping hazards by laying the cables across access ways.
AVERTISSEMENT: Protégez les chemins de câble électriques et de câbles de
télécommunication. Ne créez pas de risques de déclenchement en posant des
câbles sur les voies d'accès.