![TA DTC-300 Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/ta/dtc-300/dtc-300_getting-started-manual_4369174024.webp)
DTC-300 Getting Started Guide
Page 24
Shutting Down the DTC-300 System
Before you decide to power down your system, consider the following:
•
All of the components of your thermal analysis system are designed to be powered on for long periods.
•
The electronics of the DTC-300 perform more reliably if power fluctuations caused by turning units on
and off are minimized.
•
The over-temperature protective system is only active while the instrument is powered on.
For these reasons, turning the system and its components on and off frequently is discouraged. Therefore,
when you finish running an experiment on your instrument and wish to use the thermal analysis system for
some other task, it is recommended that you leave the instrument on.
CAUTION: Do not power off the chiller during instrument operation. The top stack, guard furnace,
and bottom stack must be below 50°C. Failure to do so may result in damage to the instrument.
To power down your system, set the DTC-300
MAIN POWER
switch to the OFF (
0
) position.
Running a DTC-300 Experiment
Sample Requirements
Test samples must be prepared with smooth and flat surfaces on both sides. Dense, rigid samples must be
tested with thermal compound applied to both surfaces, to minimize the thermal resistance of the interface
between the sample and the instrument. This might not be appropriate for porous materials, since the ther-
mal compound may penetrate the sample, thereby changing its properties. Thin, flexible samples, like
polymeric films and papers, should also be tested without thermal compound.
NOTE
: In cases when the sample needs to be tested without using thermal compound, a special calibration
must be performed, in the same conditions.
Samples may be up to 20 mm (0.75 in) thick. Unless the sample material is a product having a fixed thick-
ness, e.g. plastic film or paper, the optimum sample thickness for obtaining high test accuracy can be cal-
culated from an estimate of the sample thermal conductivity:
d < 50 *
λ
[mm]
and d > 0.5 *
λ
[mm]
In these expressions, enter for
λ
the expected thermal conductivity in W/mK.