technical reference
63
SYNRAD OEM v40 Operator’s Manual Version 4
User I/O connections
The User I/O connections section includes subsections:
■
User I/O connection summary
■
Input/output signals
■
Sample I/O circuits
User I/O connection summary
Table below provides a quick reference summary for Firestar v40 User I/O connections.
Table 4-3
User I/O pin descriptions
Pin Function
Description
1
PWM Return
Use this input pin to provide the return side of the PWM Command signal.
2
Remote Reset/Start Request
Apply a positive or negative voltage (±5–24 VDC) with respect to Pin 11, Input Common,
to disable the laser. The laser remains disabled while voltage is applied to this pin. Remov-
ing voltage from the Remote Reset/Start Request pin causes the laser’s RDY indicator to
illuminate and begins a five-second countdown after which lasing is enabled.
3
Remote Interlock
Apply a positive or negative voltage (±5–24 VDC) with respect to Pin 11, Input Common,
to enable lasing. If your system does not use remote interlock, this pin must be connected to
a voltage source in the range of ±5–24 VDC. Refer to Figure 3-7 for a diagram showing how
the Remote Interlock input is factory-jumpered.
4
+ 5 VDC Auxiliary Power
This connection pr5 VDC for driving external inputs or outputs. The +5 VDC
Auxiliary Power output can source up to 0.5 A and is protected by a 0.5 A self-resetting
fuse. The return (ground) path must be through Pin 12, Auxiliary DC Power Ground.
5
+ 24 VDC Auxiliary Power
This connection pr24 VDC for driving external inputs or outputs. The +24 VDC
Auxiliary Power output can source up to 0.5 A and is protected by a 0.5 A self-resetting
fuse. The return (ground) path must be through Pin 12, Auxiliary DC Power Ground.
6
Laser Active
This bi-directional switched output is internally connected to Pin 13, Output Common,
when the laser is actively lasing (LASE indicator illuminated red). This output is open
(high impedance) when no beam is being emitted (LASE indicator Off).
7
Over Temperature
This bi-directional switched output is internally connected to Pin 13, Output Common,
when laser temperature is above safe operating limits (TMP indicator illuminated red). The
output is open (high impedance) when laser temperature is within operating limits (TMP
indicator green).