maintenance/
troubleshooting
98
SYNRAD OEM v40 Operator’s Manual Version 4
Troubleshooting
Over temperature fault
Over temperature faults occur when thermal limits in the laser are exceeded (the TMP indica-
tor changes from green to red). Beginning with OEM D version v40 lasers, over temperature
faults are now latched. To reset an over temperature fault, lower coolant temperature below 22
°C (water-cooled) or below 40 °C (air-cooled) and then cycle DC power to the laser. Once the
RDY lamp is lit, lasing is possible after the five-second delay.
Note: Over temperature faults are not latched on earlier OEM v40 lasers. Cool the laser
until the TMP indicator changes from red to green. After the RDY lamp lights, lasing
is possible after the five-second delay.
Shutter
A closed shutter is not considered a fault condition. When a Shutter Open Request signal is
applied, lasing is enabled after a five-second delay provided that the RDY indicator is illumi-
nated yellow.
Under/over voltage fault
An under voltage or over voltage fault occurs when the DC input voltage is below or above
preset limits. This fault is indicated by the RDY indicator flashing 1 blink (under voltage) or
2 blinks (over voltage). To reset an under voltage or over voltage fault, ensure that 30 VDC
is applied to the laser under full load conditions and then cycle DC power. When the RDY
indicator illuminates, lasing is enabled after a five-second delay.
RF Drive DC fault
An RF Drive DC fault is indicated by the RDY indicator flashing 3 blinks. Reset the laser by
removing DC power to the laser, wait 30 seconds, and then reapply DC power. If the RF Drive
DC fault reappears, contact SYNRAD or a SYNRAD Authorized Distributor.
PWM Drive fault
A PWM Drive fault is indicated by the RDY indicator flashing 4 blinks. Reset the laser by
removing DC power to the laser, wait 30 seconds, and then reapply DC power. If the PWM
Drive fault reappears, contact SYNRAD or a SYNRAD Authorized Distributor.