Working with results
If you want to remove colonies, press
4. If you have chosen to add colonies, click on the image at the point you want to add the
colony – you cannot add colonies outside the frame.
If you have chosen to remove colonies, click on the colony markers you want to
remove.
The count will be updated in the Results box as you click in the image.
5. Save the change to the result (see
The count(s) will be updated in the result together with any values depending on the count(s).
An M flag will appear in the Flags column showing you have manually added or removed
colonies.
Adjusting the frame for a result
Note
You cannot adjust the frame for a Spiral Plate batch.
To adjust the frame for a Pour Plate result:
1. Select the Edit Result tab for the plate result – see
2. Press
3. Adjust the frame using the same techniques as when you create a batch or before taking
a measurement – see
Adjusting the position and size of frames
4. Press
again.
The plate will be remeasured using the adjusted frame.
5. Save the change to the result (see
Rejecting small particles in a result
When you create a batch, you can adjust the setting of the Reject Small Particles filter to
remove the effect of small colored regions of the image that may be wrongly interpreted as
colonies – see
, page 30. You can also change the Reject Small Particles
filter in Measurement Mode
before
taking a measurement (see
This section shows you how to change the Reject Small Particles setting for an individual
plate
after
it has been measured.
84
aCOLyte3
User Manual
Содержание SYNBIOSIS aCOLyte3
Страница 2: ......
Страница 6: ......
Страница 12: ...Getting started 6 aCOLyte3 User Manual...
Страница 22: ...Capturing images 16 aCOLyte3 User Manual...
Страница 66: ...Creating a new batch 60 aCOLyte3 User Manual...
Страница 96: ...Working with results 90 aCOLyte3 User Manual...
Страница 108: ...Configuring aCOLyte3 102 aCOLyte3 User Manual...