VelaShape III User Manual Troubleshooting
PB81499EN
79
Table 4: System Errors Messages Displayed On Screen and/or Applicator Panel
Message
Symptom
Corrective Action
Improve Vacuum
Seal
Vacuum is not implemented
correctly upon the skin.
●
Remove and clean the
applicator cover.
●
Replace the cover filter.
●
Put on a new cover.
●
Make sure the patient skin is
cleaned and shaved.
RF Out of Range
The system cannot apply RF
correctly.
●
Make sure there is a good
vacuum seal between the
applicator cover and patient
skin,
●
Make sure there is an
adequate amount of gel.
●
Make sure the applicator is
placed on the skin
appropriately.
No applicator, No
Tip
System freezes, no RF or IR
pulses emitted to the patient's
skin even if footswitch/trigger
is pressed.
●
Let go of footswitch/trigger and
make sure that the applicator
and tip are well connected and
sealed.
●
Let go of footswitch/trigger and
wait 10 minutes for the system
to recognize the applicator and
tip.
●
If after 10 minutes the system
does not recognize the
applicator and tip restart the
system.
●
If after restart the system does
not recognize the applicator
and tip unplug other medical
devices in the room and restart
the system again.