VelaShape III User Manual Cellulite and Circumference Treatment
PB81499EN
59
Treatment Parameters
Applicator
Cover
Cycle
Step
Parameter
Level
VContour
Large
Cycle #1
Step #1
Vacuum
Off
RF
3
(
Step #2
Vacuum
Off
RF
3 / 2
(
2
)
Step #3
Vacuum
Off
RF
3 / 2
(
2
)
Medium
Cycle #2
Step #1
Vacuum
1 – 3
(
RF
3 / 2
(
2
)
Step #2
Vacuum
1 – 3
(
3
)
RF
3 / 2
(
2
)
Step #3
Vacuum
1 – 3
(
3
)
RF
3 / 2
(
2
)
If the patient feels uncomfortable you may:
1.
Re-apply VelaSpray Ease lotion at minimum intervals of one minute, in
each step of each cycle.
2.
Use the "
Quick Technique
" manner
when stacking one pulse.
3.
Only after trying both options 1 & 2 above and the patient still feels
uncomfortable, reduce the RF level to 2.
e
Maintain RF Level at 3 whenever possible to achieve an optimized result.
f
Option to increase vacuum from 1 -> 2 -> 3 based on patient tolerance (in areas where there is
minimal laxity).
g
"
Quick Technique
" manner: stacking one pulse in each segment, while moving the applicator
faster in the “
Horizontal snake-like
” or “
Vertical snake-like
” manner according to the step.