System Installation
PB81499EN
30
Moving the System within the Facility
1.
Turn the system off and disconnect the power cable.
2.
Disconnect the applicator and place it in a safe location.
3.
Release the wheel brakes at the base of the system.
4.
Slowly push or pull the system using the system's handle.
Moving the System to another Facility
1.
Turn the system off and disconnect the power cable.
2.
Disconnect the applicator and pack it in its storage case.
3.
Release the wheel brakes at the base of the system.
4.
Slowly push or pull the system using the system's handle and carefully lift it up
onto the vehicle. Spread padding on the wall of the transport vehicle to avoid
scratching the system's panels and strap the system to the wall.
5.
Upon arrival at the new facility, position the system in its new location, using
the system handle.
6.
Reconnect the applicator, footswitch and the power cable.
Caution
Do not
lay the system on its back or sides! Doing so may damage the
vacuum pump.
Electrical Requirements
The system incorporates a universal power supply module that allows it to be
connected to standard electrical outlets anywhere. The system's electrical
requirements are:
●
100-230 VAC; up to 4A; 50-60 Hz; single phase
The electrical requirements are printed on the system's nameplate, located on the
system's rear panel.
Input power lines should be free of transients, voltage and current spikes, sags
and surges. Consequently, the system power line should not be shared with other
heavy variable loads such as elevators, air conditioning systems, large motors,
etc.