2. Service
Functional-Group-Diagnostic Trees (Diagnostic Section 2)
73
2.1.5.1.0 Internal-Battery Charger Test
If the 5071A operates for less time than specified, the fault is either the
battery or that it is not being charged properly. Verify charger operation as
follows:
1
Apply ac power and observe the instrument complete warm-up.
2
Remove ac power and do not disconnect or turn-off internal-standby
batteries.
3
Allow the instrument to run on the internal-standby batteries for at least one
minute.
4
Reapply ac power to the instrument. (This causes the boost-charge mode
entry: the current to the batteries should be between 140 mA and 180 mA., if
the charger is operating correctly.
5
Verify that the boost-charge current is between 140 and 180 mA using a clip-
on ammeter at any battery lead.
QUESTION:
Is the boost-charge current within specification?
If No:
Remove power, replace A11, and retest charging current.
If Yes:
Replace both battery packs and retest operating time.
2.1.5.2.0 A11 F1 Fuse Fault
If the instrument will not operate from the internal batteries at all, the fault
could be fuse F1 on the A11 board. Fuse F1 is in series with the batteries to
protect them from catastrophic discharge if a short circuit occurs.
1
Ensure that both batteries are connected in the normal configuration and all
external instrument power is removed,
2
Measure the voltage at TP1 on A11 with respect to a common point (not
chassis ground) such as TP's 6,9, and 12.
3
Verify a battery voltage of approximately 24 volts.
QUESTION:
Is 24 volts present?
If Yes:
The battery packs may be at fault or are not being properly
charged. Go to, 2.1.5.1.0.
If No:
Disconnect the two internal-standby battery connections and
replace F1 on A11 (or A11 entirely). Reconnect the internal-
standby battery connections and retest. If only F1 is replaced
and it blows again at retest, replace A11 entirely, then retest.
Содержание 5071A
Страница 2: ......
Страница 40: ...1 Performance Tests Verifying Specifications 5071A Complete Performance Tests 24 ...
Страница 48: ...2 Service Pre Troubleshooting Information 32 Figure 2 1 Instrument Top View ...
Страница 147: ...3 Theory of Operation Functional Block Descriptions 131 Figure 3 2 5071A Overall Block Diagram ...
Страница 148: ...3 Theory of Operation Functional Block Descriptions 132 Figure 3 2 5071A Overall Block Diagram Continued ...
Страница 164: ...3 Theory of Operation Functional Block Descriptions 148 Figure 3 3 A1 Power Supply Distribution Schematic Diagram ...
Страница 166: ...3 Theory of Operation Functional Block Descriptions 150 Figure 3 4 A1 Slot Module Interconnect Diagram ...
Страница 168: ...3 Theory of Operation Functional Block Descriptions 152 Figure 3 5 Power Supply Block Diagram ...
Страница 171: ...3 Theory of Operation Functional Block Descriptions 155 Figure 3 7 Opt 048 A1 Slot Module Interconnect Diagram ...
Страница 172: ...3 Theory of Operation Functional Block Descriptions 156 Figure 3 8 Opt 048 Power Supply Block Diagram ...
Страница 173: ......
Страница 175: ...3 Theory of Operation Functional Block Descriptions 159 Figure 3 9 A1 Component Locator Diagram ...
Страница 194: ...5 Replaceable Parts How To Order A Part 178 ...
Страница 197: ......