![Symmetricom 5071A Скачать руководство пользователя страница 116](http://html.mh-extra.com/html/symmetricom/5071a/5071a_assembly-and-service-manual_1419613116.webp)
2. Service
Assembly/Module Diagnostic Trees (Diagnostic Section 3)
100
3.4.2.1 A19 10-MHz Output Test
1
Disconnect cable from A19 output connector.
2
Connect A19 output connector to a scope with 50
Ω
input impedance (If the
scope doesn't have a 50
Ω
input, use the high impedance input with a 50
Ω
coaxial feedthrough terminator installed at the scope end of the coax).
3
Verify presence of a 10-MHz undistorted sine wave with at least
1 Vp-p amplitude.
QUESTION:
Does the scope show a 10-MHz undistorted sine wave
with at least 1 Vp-p amplitude?
If Yes:
Remove power, replace cable between A9J2 and A10J5,
and
retest.
If No:
Go to, 3.9.1.1 (A19 trouble tree).
3.4.3.0 A9 80-MHz Signal Test
1
Connect A9J2 to the 50
Ω
input of a scope using coax. (If a 50
Ω
input is
absent, use the high impedance input with a 50
Ω
coaxial feedthrough
terminator at the scope end of the coax).
2
Verify presence of an 80-MHz undistorted sine wave with at least
0.6 Vp-p amplitude.
3
Repeat steps 1 and 2 for A9J3 and A9J4
QUESTION:
Does the scope show an 80-MHz. undistorted sine wave
with at least 0.6 Vp-p amplitude present at all three test
connections?
If Yes:
Go to, 3.4.4.0.
If No:
Remove power, replace A9, and retest.
3.4.4.0 A9 J1 320-MHz Signal Test
1
Adjust input attenuation on spectrum analyzer for a +30 dBm maximum
signal.
2
Connect A9J1 to the spectrum analyzer using coax.
3
Verify presence of a 320-MHz signal of at least +18 dBm amplitude.
QUESTION:
Does the analyzer show a 320-MHz signal of at least
+18 dBm amplitude?
If Yes:
Go to, 3.4.5.0.
If No:
Remove power, replace A9, and retest.
Содержание 5071A
Страница 2: ......
Страница 40: ...1 Performance Tests Verifying Specifications 5071A Complete Performance Tests 24 ...
Страница 48: ...2 Service Pre Troubleshooting Information 32 Figure 2 1 Instrument Top View ...
Страница 147: ...3 Theory of Operation Functional Block Descriptions 131 Figure 3 2 5071A Overall Block Diagram ...
Страница 148: ...3 Theory of Operation Functional Block Descriptions 132 Figure 3 2 5071A Overall Block Diagram Continued ...
Страница 164: ...3 Theory of Operation Functional Block Descriptions 148 Figure 3 3 A1 Power Supply Distribution Schematic Diagram ...
Страница 166: ...3 Theory of Operation Functional Block Descriptions 150 Figure 3 4 A1 Slot Module Interconnect Diagram ...
Страница 168: ...3 Theory of Operation Functional Block Descriptions 152 Figure 3 5 Power Supply Block Diagram ...
Страница 171: ...3 Theory of Operation Functional Block Descriptions 155 Figure 3 7 Opt 048 A1 Slot Module Interconnect Diagram ...
Страница 172: ...3 Theory of Operation Functional Block Descriptions 156 Figure 3 8 Opt 048 Power Supply Block Diagram ...
Страница 173: ......
Страница 175: ...3 Theory of Operation Functional Block Descriptions 159 Figure 3 9 A1 Component Locator Diagram ...
Страница 194: ...5 Replaceable Parts How To Order A Part 178 ...
Страница 197: ......