2. Service
Top-Level Diagnostic Tree (Diagnostic Section 1)
50
FIRST WORD (Rightmost word):
-
0 0 0 0 0 0 0 0
⇐
Hex digit
/ \ / \ / \ / \ / \ / \ / \ / \
00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00
⇐
Binary digit
| | | | | | | | | | | | | | | \_1. Servo offset - No Message
| | | | | | | | | | | | | | \_2. CBT shutdown: <#> - Fatal Error
| | | | | | | | | | | | | \_3. VCXO near tuning limit - Warning
| | | | | | | | | | | | \_4. VCXO at tuning limit - Fatal Error
| | | | | | | | | | | \_5. C-field current at range limit - Fatal Error
| | | | | | | | | | \_6. E-multiplier voltage at range limit - Message
| | | | | | | | | \_7. Signal gain at range limit - Warning
| | | | | | | | \_8. Internal DC supply out of range - Warning
| | | | | | | \_9. Mass spec supply regulation failure - Fatal Error
| | | | | | \_10. Hotwire supply regulation failure - Fatal Error
| | | | | \_11. CBT oven supply regulation failure - Fatal Error
| | | | \_12. CBT oven voltage at range limit - Fatal Error
| | | \_13. RF amplitude at range limit - Warning
| | \_14. Microprocessor PLL unlocked - Warning
| \_15. 87 MHz PLL unlocked - Message
\_16. Unused bit - always zero.
SECOND WORD:
0 0 0 0 0 0 0 0
⇐
Hex digit
/ \ / \ / \ / \ / \ / \ / \ / \
00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00
⇐
Binary digit
| | | | | | | | \
_17. DRO unlocked - Message
| | | | | | | \
_18. VCXO oven error - Message
| | | | | | \
_19. CBT signal loss - Fatal Error
| | | | | \
_20. CBT signal low - Warning
| | | | \
_21. Servo system is in manual mode - Warning
| | | \
_22. Attempted lock on wrong peak - Fatal Error
| | \
_23. Center peak saturated - Message
\-----
reserved
--------/
Example Error-Code Interpretation
-
You can create a fatal error by disconnecting A5P2 from A5 after the instrument has
powered-up and is running (wait up to 35 seconds for the software to confirm this
failure). When the attention LED begins flashing, you're ready to read the error code
from the log.
-
Typical log display:
Error code: 000000a0 20000000
-
By using the above error-message map, the following information may be decoded:
-
15. 87 MHz PLL unlocked - Message
-
19. CBT signal loss - Fatal Error (This is the event that initiated the shut-down.)
-
20. CBT signal low – Warning
NOTE
You must power-off, reconnect A5P2, and then power-on again
to clear this error
.
Содержание 5071A
Страница 2: ......
Страница 40: ...1 Performance Tests Verifying Specifications 5071A Complete Performance Tests 24 ...
Страница 48: ...2 Service Pre Troubleshooting Information 32 Figure 2 1 Instrument Top View ...
Страница 147: ...3 Theory of Operation Functional Block Descriptions 131 Figure 3 2 5071A Overall Block Diagram ...
Страница 148: ...3 Theory of Operation Functional Block Descriptions 132 Figure 3 2 5071A Overall Block Diagram Continued ...
Страница 164: ...3 Theory of Operation Functional Block Descriptions 148 Figure 3 3 A1 Power Supply Distribution Schematic Diagram ...
Страница 166: ...3 Theory of Operation Functional Block Descriptions 150 Figure 3 4 A1 Slot Module Interconnect Diagram ...
Страница 168: ...3 Theory of Operation Functional Block Descriptions 152 Figure 3 5 Power Supply Block Diagram ...
Страница 171: ...3 Theory of Operation Functional Block Descriptions 155 Figure 3 7 Opt 048 A1 Slot Module Interconnect Diagram ...
Страница 172: ...3 Theory of Operation Functional Block Descriptions 156 Figure 3 8 Opt 048 Power Supply Block Diagram ...
Страница 173: ......
Страница 175: ...3 Theory of Operation Functional Block Descriptions 159 Figure 3 9 A1 Component Locator Diagram ...
Страница 194: ...5 Replaceable Parts How To Order A Part 178 ...
Страница 197: ......