2. Service
Top-Level Diagnostic Tree (Diagnostic Section 1)
65
Advisory Messages
(Subsection 7)
Advisory Messages are found in the log. If they are associated with fatal
errors or warnings, they may be useful in fault isolation.
1.7.1 Advisory Message Interpretation
Message: E-mult adjusted - this appears to notify that due to internal CBT
parameters the E-mult voltage was adjusted by a larger amount than usually
required during steady-state operation. The appearance of this message on
the display will be accompanied by a visible "wink" of the green Continuous
Operation LED which is normal. No action required.
Message:
E-multiplier at voltage range limit
. This message may
occur with CBT aging. The condition does not by itself imply any loss of
performance.
No action required.
Message:
87 MHz PLL unlocked
. This message indicates an RF-chain
problem leading to an erroneous voltage level in the A5 module. This
message may be associated with a fatal error.
Go to, 1.3.7.
Message:
DRO unlocked
. This message indicates an RF-chain problem
leading to an erroneous voltage level in the A15 module. This message may
be associated with a fatal error.
Go to, 1.3.7.
Message:
VCXO Oven error
. This message indicates a problem with the
A19 module or its monitoring circuit. A malfunction in the A19 module will
usually cause degraded performance or a fatal error.
Go to, 3.9.1.1.
Message:
Center peak saturated
. This message indicates that the
amplified CBT output signal is unstable. If the condition persists, a CBT
signal loss fatal error may result.
Check the analog signal chain (2.2.1.0), and the Electron Multiplier supply
(3.7.2.1).
If no problems are found with A6, A14, or A16 with the above procedures, the
CBT may be faulty. Contact Symmetricom for further assistance.
Содержание 5071A
Страница 2: ......
Страница 40: ...1 Performance Tests Verifying Specifications 5071A Complete Performance Tests 24 ...
Страница 48: ...2 Service Pre Troubleshooting Information 32 Figure 2 1 Instrument Top View ...
Страница 147: ...3 Theory of Operation Functional Block Descriptions 131 Figure 3 2 5071A Overall Block Diagram ...
Страница 148: ...3 Theory of Operation Functional Block Descriptions 132 Figure 3 2 5071A Overall Block Diagram Continued ...
Страница 164: ...3 Theory of Operation Functional Block Descriptions 148 Figure 3 3 A1 Power Supply Distribution Schematic Diagram ...
Страница 166: ...3 Theory of Operation Functional Block Descriptions 150 Figure 3 4 A1 Slot Module Interconnect Diagram ...
Страница 168: ...3 Theory of Operation Functional Block Descriptions 152 Figure 3 5 Power Supply Block Diagram ...
Страница 171: ...3 Theory of Operation Functional Block Descriptions 155 Figure 3 7 Opt 048 A1 Slot Module Interconnect Diagram ...
Страница 172: ...3 Theory of Operation Functional Block Descriptions 156 Figure 3 8 Opt 048 Power Supply Block Diagram ...
Страница 173: ......
Страница 175: ...3 Theory of Operation Functional Block Descriptions 159 Figure 3 9 A1 Component Locator Diagram ...
Страница 194: ...5 Replaceable Parts How To Order A Part 178 ...
Страница 197: ......