Safety Considerations and Warranty
v
Safety Considerations (contd)
WARNING:
FOR CONTINUED PROTECTION AGAINST FIRE, REPLACE THE LINE FUSE(S)
ONLY WITH 250V FUSE(S) OF THE SAME CURRENT RATING AND TYPE (FOR
EXAMPLE, NORMAL BLOW, TIME DELAY). DO NOT USE REPAIRED FUSES OR
SHORT CIRCUITED FUSEHOLDERS.
AVERTISSEMENT:
POUR LA PROTECTION CONTINUÉE CONTRE LE FEU, REMPLACER LE
FUSIBLE DE LIGNE (LES FUSIBLES) SEULEMENT AVEC 250V FUSIBLE (LES
FUSIBLES) DU MÊME COURANT ÉVALUANT ET TAPE (PAR EXEMPLE, LE
COUP NORMAL, LE DÉLAI). NE PAS UTILISER DE FUSIBLES RÉPARÉS OU
CIRCUITED FUSEHOLDERS COURT.
---------------------------------------------------------------------------------------------------------
DC Power Supply (VDC Operation Model) Installation
•
Use a 15 Amp DC circuit breaker in series with the DC power source.
•
Do not connect the unit directly to a DC power source without the breaker.
•
14 AWG (1.5mm
2
) gage wire is the minimum gage permitted by the NEC for DC
power source hookup.
•
The Unit Chassis must be grounded for proper safety.
---------------------------------------------------------------------------------------------------------
WARNING:
ENSURE THAT A DISCONNECT DEVICE, SUCH AS A SWITCH, WITH THE
APPROPRIATE VOLTAGE/CURRENT RATING IS PROVIDED WHEN CONNECTING A
DC POWER SOURCE TO THE VDC OPERATION MODEL.
AVERTISSEMENT:
S'ASSURER QU'UN DÉBRANCHER L'APPAREIL, TEL QU'UN COMMUTATEUR, AVEC
LE CLASSEMENT DE TENSION/COURANT APPROPRIÉ EST FOURNI EN
CONNECTANT UNE SOURCE DE POUVOIR DE DC AU MODÈLE DE
CONFIGURATION DE VDC
.
--------------------------------------------------------------------------------------------------------
Acoustic Noise Emissions
LpA<47 dB at operator position, at normal operation, tested per EN 27779. All data are the results
from type test.
Geräuschemission
LpA<47 dB am Arbeits platz, normaler Betrieb, geprüft nach EN 27779. Die Angagen beruhen auf
Ergebnissen von Typenprüfungen.
Electrostatic Discharge Immunity Testing
When the product is tested with 8kV AD, 4kV CD and 4kV ID according to IEC801-2, a system
error
may occur that may affect measurement data made during these disturbances. After these
occurrences, the system self-recovers without user intervention.
Содержание 5071A
Страница 2: ......
Страница 40: ...1 Performance Tests Verifying Specifications 5071A Complete Performance Tests 24 ...
Страница 48: ...2 Service Pre Troubleshooting Information 32 Figure 2 1 Instrument Top View ...
Страница 147: ...3 Theory of Operation Functional Block Descriptions 131 Figure 3 2 5071A Overall Block Diagram ...
Страница 148: ...3 Theory of Operation Functional Block Descriptions 132 Figure 3 2 5071A Overall Block Diagram Continued ...
Страница 164: ...3 Theory of Operation Functional Block Descriptions 148 Figure 3 3 A1 Power Supply Distribution Schematic Diagram ...
Страница 166: ...3 Theory of Operation Functional Block Descriptions 150 Figure 3 4 A1 Slot Module Interconnect Diagram ...
Страница 168: ...3 Theory of Operation Functional Block Descriptions 152 Figure 3 5 Power Supply Block Diagram ...
Страница 171: ...3 Theory of Operation Functional Block Descriptions 155 Figure 3 7 Opt 048 A1 Slot Module Interconnect Diagram ...
Страница 172: ...3 Theory of Operation Functional Block Descriptions 156 Figure 3 8 Opt 048 Power Supply Block Diagram ...
Страница 173: ......
Страница 175: ...3 Theory of Operation Functional Block Descriptions 159 Figure 3 9 A1 Component Locator Diagram ...
Страница 194: ...5 Replaceable Parts How To Order A Part 178 ...
Страница 197: ......