2. Service
Top-Level Diagnostic Tree (Diagnostic Section 1)
52
1.3.2.1 Cs Oven Failed: Ion Pump Overcurrent Warm-up Error
This error occurs when current drawn by the CBT Ion Pump fails to go
below 40
μ
A within 45 minutes after operation begins. It can be
caused by either excessive current drawn by the CBT as a result of a
high internal gas pressure, or by a failure in the Ion pump high-voltage
supply (P/O A16), or the current-monitoring circuit (P/O A2). The
following procedures check these alternatives:
1
Observe the measured Ion-Pump supply current on the LCD display
(Use the
INFO/Pump
menu selection.).
QUESTION:
Is the current more than 40
μ
A?
If Yes:
Go to, 1.3.2.2.
If No:
The problem may have cleared itself, remove ac power,
and retest.
1.3.2.2 A16/CBT Excess Ion-Pump Current Test
1
Remove instrument power and wait at least 60 seconds for the high voltage
supplies to bleed down.
2
Disconnect the Ion Pump high-voltage (HV) connector from the CBT (see
Figure 2.1). (Ensure that the end of the high-voltage lead is clear of any
conductors.)
3
Disconnect the flexible cable joining the CBT to A2J2.
4
Power up the instrument and allow a few seconds for it to enter the Fatal-
Error mode.
5
Observe the measured Ion Pump supply current on the LCD display (Use the
INFO/Pump
menu selection.).
6
Remove power, wait 1 minute, and reconnect the Ion-pump HV lead.
QUESTION:
Was the current more than 1 microamp?
If Yes:
A fault is present in the Ion Pump power supply circuitry,
Go to, 3.7.1.0, A16 diagnostic tree (This tree also checks part of
A2.).
If No:
CBT Ion-Pump current is high. When caused by poor vacuum (for
example, due to prolonged storage) the condition may correct
itself if the unit is placed into
STANDBY
mode for 24 hours. If the
condition persists after this time, remove power and contact
Symmetricom for assistance.
Содержание 5071A
Страница 2: ......
Страница 40: ...1 Performance Tests Verifying Specifications 5071A Complete Performance Tests 24 ...
Страница 48: ...2 Service Pre Troubleshooting Information 32 Figure 2 1 Instrument Top View ...
Страница 147: ...3 Theory of Operation Functional Block Descriptions 131 Figure 3 2 5071A Overall Block Diagram ...
Страница 148: ...3 Theory of Operation Functional Block Descriptions 132 Figure 3 2 5071A Overall Block Diagram Continued ...
Страница 164: ...3 Theory of Operation Functional Block Descriptions 148 Figure 3 3 A1 Power Supply Distribution Schematic Diagram ...
Страница 166: ...3 Theory of Operation Functional Block Descriptions 150 Figure 3 4 A1 Slot Module Interconnect Diagram ...
Страница 168: ...3 Theory of Operation Functional Block Descriptions 152 Figure 3 5 Power Supply Block Diagram ...
Страница 171: ...3 Theory of Operation Functional Block Descriptions 155 Figure 3 7 Opt 048 A1 Slot Module Interconnect Diagram ...
Страница 172: ...3 Theory of Operation Functional Block Descriptions 156 Figure 3 8 Opt 048 Power Supply Block Diagram ...
Страница 173: ......
Страница 175: ...3 Theory of Operation Functional Block Descriptions 159 Figure 3 9 A1 Component Locator Diagram ...
Страница 194: ...5 Replaceable Parts How To Order A Part 178 ...
Страница 197: ......