AMU Rescon
Default Values
A-96.250.351 / 230320
65
Scaling: Range low:................................................................ 0.0 °C
Scaling: Range high: ............................................................50.0 °C
Alarm Relay:
Alarm:
Resistivity
Alarm high: .......................................................................... 200 MΩ
Alarm low:........................................................................... 0.00 MΩ
Hysteresis:.......................................................................... 1.00 MΩ
Conductivity
Alarm high: ..........................................................................2000
µ
S
Alarm low:.......................................................................... 0.000
µ
S
Hysteresis:..........................................................................10.00
µ
S
Delay: .......................................................................................... 5 s
Sample Flow; Flow Alarm........................................................... yes
Sample Flow; Alarm high.......................................................120 l/h
Sample Flow; Alarm low ............................................................5 l/h
Sample Temp; Alarm high:....................................................... 90 °C
Sample Temp; Alarm low: ..........................................................0 °C
Case temp. high: ..................................................................... 65 °C
Case temp. low:.........................................................................0 °C
Relay 1 and 2
Function:.......................................................................... limit upper
Parameter:.................................................................... Meas. Value
Resistivity
Setpoint: .............................................................................. 200 MΩ
Hysteresis:............................................................................... 1 MΩ
Conductivity
Setpoint: ..............................................................................1000
µ
S
Hysteresis:..........................................................................10.00
µ
S
Delay: ........................................................................................ 30 s
If Function = Control upw. or dnw:
Parameter:
....................................................................Meas. Value
Settings: Actuator: .......................................................... Frequency
Settings: Pulse Frequency: ..............................................120/min
Resistivity
Settings: Control Parameters: Setpoint:........................... 200 MΩ
Settings: Control Parameters: P-band: ................................ 1 MΩ
Conductivity
Settings: Control Parameters: Setpoint:...........................1000
µ
S
Settings: Control Parameters: P-band: ...........................10.00
µ
S
Parameter:
...................................................................Temperature
Settings: Actuator: .......................................................... Frequency
Settings: Pulse Frequency: ..............................................120/min
Settings: Control Parameters: Setpoint:...............................50 °C
Settings: Control Parameters: P-band: ..................................1 °C
Parameter:
................................................................... Sample flow
Settings: Actuator: .......................................................... Frequency