40
A-96.250.351 / 230320
AMU Rescon
Error List
Error
Description
Corrective action
E001
Alarm high
– check process
– check programmed value
E002
Alarm low
– check process
– check programmed value
E007
Sample Temp. high
– check process
– check programmed value
E008
Sample Temp. low
– check process
– check programmed value
E009
Sample Flow high
– check sample flow
– check programmed value
E010
Sample Flow low
– establish sample flow
– clean instrument
– check programmed value
E011
Temp. shorted
– Check wiring of sensor
– Check sensor
E012
Temp. disconnected
– Check wiring of sensor
– Check sensor
E013
Case Temp. high
– check case/environment temperature
– check programmed value
E014
Case Temp. low
– check case/environment temperature
– check programmed value
E015
USP Error
– check process
E017
Control Timeout
– check control device or programming in
Installation, Relay contact, Relay 1/2
E019
Sensor shorted
– Check wiring of sensor
– Check sensor
E020
Sensor interrupted
– Check wiring of sensor
– Check sensor
E024
Input active