background image

AMU Rescon

Program Overview

A-96.250.351 / 230320

43

8.2. Diagnostics (Main Menu 2)

Identification

Designation

AMU Rescon

* Menu numbers

 2.1*

Version

V6.20-11/16

Factory Test

Instrument

2.1.3.1*

2.1.3*

Motherboard
Front End

Operating Time

Years / Days / Hours / Minutes / Seconds

2.1.4.1*

2.1.4*

Sensors

Cond. Sensor

Current Value MOhm

 2.2*

2.2.1*

(Raw value) MOhm
Cell Constant

Cal. History

Number

2.2.1.5.1*

2.2.1.5*

Date, Time
RSlo (KOhm)

Miscellaneous

Case Temp.

2.2.2.1*

2.2.2*

Sample

Sample ID

2.3.1*

 2.3*

Temperature °C
Nt5K Ohm

I/O State

Alarm Relay

2.4.1*

2.4*

Relay 1/2

2.4.2*

Input
Signal Output 1/2

Interface

Protocol

2.5.1*

2.5*

Baud rate

Содержание AMU Rescon

Страница 1: ...AMU Rescon Version 6 20 and higher A 96 250 351 230320 Operator s Manual ...

Страница 2: ...nical specialists around the world For any technical question contact your nearest SWAN representative or the manufacturer SWAN ANALYTISCHE INSTRUMENTE AG Studbachstrasse 13 8340 Hinwil Switzerland Internet www swan ch E mail support swan ch Document Status Title AMU Rescon Operator s Manual ID A 96 250 351 Revision Issue 00 March 2007 First Edition 01 June 2019 Update to firmware V6 20 ...

Страница 3: ...Installation 14 3 1 Installation Checklist Monitors 14 3 2 Dimensions of the AMU Transmitter 15 3 3 Connecting Sample Inlet and Outlet 16 3 3 1 Sample Inlet 16 3 3 2 Sample Outlet 16 3 4 Install Sensor RCU 17 3 5 Electrical Connection 19 3 6 Power supply 21 3 7 Sensor 21 3 8 Flow Meter 21 3 9 Input 21 3 10 Relay Contacts 22 3 10 1 Alarm Relay 22 3 10 2 Relay 1 and 2 22 3 11 Signal Output 1 and 2 c...

Страница 4: ...esistivity QC Kit Test Plug 35 6 4 1 Introduction 35 6 4 2 Carry out a transmitter check 36 6 5 Fine Adjust 37 6 6 Longer Stop of Operation 38 7 Error List 39 8 Program Overview 42 8 1 Messages Main Menu 1 42 8 2 Diagnostics Main Menu 2 43 8 3 Maintenance Main Menu 3 44 8 4 Operation Main Menu 4 44 8 5 Installation Main Menu 5 45 9 Program List and Explanations 47 1 Messages 47 2 Diagnostics 47 3 ...

Страница 5: ...instructions are given throughout this manual at the respective locations where observation is most important Strictly follow all safety instructions in this publication Target audience Operator Qualified person who uses the equipment for its intended purpose Instrument operation requires thorough knowledge of applications instrument functions and software program as well as all applicable safety ...

Страница 6: ...the prevention instructions carefully WARNING Severe injuries or damage to the equipment can occur if such warnings are ignored Follow the prevention instructions carefully CAUTION Damage to the equipment minor injury malfunctions or incor rect process can be the consequence if such warnings are ig nored Follow the prevention instructions carefully Mandatory Signs The importance of the mandatory s...

Страница 7: ...ponsible for proper system operation All precautions must be followed to ensure safe operation of the in strument Spare Parts and Disposables Use only official SWAN spare parts and disposables If other parts are used during the normal warranty period the manufacturer s warranty is voided Electrical shock hazard Corrosive Harmful to health Flammable Warning general Attention general ...

Страница 8: ...ormed by authorized personnel only Whenever electronic service is required disconnect instru ment power and power of devices connected to relay 1 relay 2 alarm relay WARNING For safe instrument installation and operation you must read and understand the instructions in this manual WARNING Only SWAN trained and authorized personnel shall perform the tasks described in this document 1 3 Restrictions...

Страница 9: ...nate these effects the temperature is determined simultaneously with an incorporated NT5K temperature sensor Several temperature compensation curves for different applications are available Signal Outputs Two signal outputs programmable for measured values freely scal able linear bilinear log or as continuous control output control parameters programmable Current loop 0 4 20 mA Maximal burden 510 ...

Страница 10: ...conductivity or resistivity standard measurement with automatic temperature compensation is performed If the USP Operating Mode ils set to on the temperature com pensation function is disabled The uncompensated measuring val ues are compared with the values of an implemented table defined by USP see table below If the deviation of the measured values is too high the Error 15 USP Error is issued Co...

Страница 11: ...mple inlet F It flows through the flow regulating valve E where the flow rate can be adjusted Then the sample flows through the flow cell block D were the resistivity of the sample is measured The sample leaves the flow cell block via flow meter through the sample outlet C A B C Sensor RC U Flow sensor Sample outlet D E F Flow cell block Flow regulating valve Sample inlet A B C D E F ...

Страница 12: ...5 x 45 mm 96 x 96 x 120 mm DIN 43700 0 45 kg Power supply Voltage Power consumption 100 240 VAC 10 50 60 Hz 5 or 24 VDC 15 max 8 VA Sensor type Two electrode sensor with built in NTC temperature probe e g Swansensor RCU Measuring range Measuring range 0 001 to 200 00 MΩ cm Resolution 0 01 MΩ cm 0 005 to 2 999 µS cm 3 00 to 29 99 µS cm 30 0 to 99 9 µS cm 100 to 1000 µS cm 0 001 µS cm 0 01 µS cm 0 1...

Страница 13: ...ual flow regulating valve Con nection to tube with Swagelok adapter For one sensor with NPT thread 33 38 44 139 28 135 151 234 96 10 45 60 75 Technical data Sample inlet Sample outlet Sample temperature Sample flow QV Flow Sample flow QV HFlow Sample inlet pressure Sample outlet pressure Swagelok G 1 4 thread Serto elbow for 6 mm flexible tube 0 60 C 3 25 l h 10 120 l h max 15 bar at 50 C pressure...

Страница 14: ... made of stainless steel SS316L to connect to tubes For one sensor with NPT thread fitting length 89 mm 15 26 33 40 7 17 130 214 96 10 45 60 75 Technical data Sample inlet Sample outlet Sample temperature Sample inlet pressure Swagelok G 1 8 thread Swagelok G 1 8 thread 10 130 C max 10 bar at 130 C ...

Страница 15: ... resistivity in high purity water Swansensor RC U with integrated cable Swansensor RC U with Plug 28 156 89 231 89 Specifications RC U Sensor Measuring range Operating temperature Pressure Accuracy at 25 C Temperature sensor Thread 0 055 1000 S cm 0 01 200 M 10 to 90 C max 10 bar at 90 C 0 5 up to 20 S cm 1 from 20 s cm to 1000 S cm NT5K 3 4 NPT ...

Страница 16: ...t sen sors Electrical Connection p 19 Power up Establish Sample Flow p 25 Switch on power First the instrument performs a self test displays the firmware version and then starts normal operation Instrument set up Program all necessary parameters see Programming p 25 Program the flow cell type Program the measuring mode resistivity or conductivity Program all sensor data cell constant temperature c...

Страница 17: ...AMU Rescon Installation A 96 250 351 230320 15 3 2 Dimensions of the AMU Transmitter 96 mm 89 8 mm 89 8 mm 112 mm 7 mm 96 mm 92 mm 92 mm ...

Страница 18: ...tallation 1 Insert the compression ferrule C and the compression cone D into the union nut B 2 Screw on the union nut onto the body do not tighten it 3 Push the stainless steel pipe through the union nut as far as it reaches the stop of the body 4 Mark the union nut at 6 o clock position 5 While holding the fitting body steady tighten the nut union 1 rotation using an open ended spanner 3 3 2 Samp...

Страница 19: ... proceed as follows Install Sensor into Flow Cell Install Sensor into a tube Installation in a by pass is recommended Choose the installation point carefully to ensure the sensor is always filled with water even if sample flow is interrupted To avoid the formation of air bubbles inside the sensor make sure that the air vent holes are always sub merged A B C Teflon tape Sensor thread Flow cell A B ...

Страница 20: ... Wrap 7 turns of teflon tape around the sensor thread 2 Screw the sensor into the flow cell or the pipe flange 3 Tighten the sensor well with a monkey spanner A B C Sensor Flange Air vent hole D E Tube Flow direction A B C D E Sensor Flange Air vent hole Flow direction Tube A B C D E A B C E D ...

Страница 21: ...0320 19 3 5 Electrical Connection CAUTION Use only the terminals shown in this diagram and only for the mentioned purpose Use of any other terminals will cause short circuits with possible corresponding consequences to material and personnel ...

Страница 22: ...20 A 96 250 351 230320 AMU Rescon Installation Rear view of AMU Transmitter 1 L RS232 N 2 3 4 5 6 7 8 9 RON OFF A PB B PB 10 11 12 13 14 24 23 22 21 20 19 ...

Страница 23: ...near the instrument easily accessible to the operator marked as interrupter for AMU Rescon 3 7 Sensor Connect the sensor to the AMU transmitter according to the con nection diagram see Electrical Connection p 19 For sensor settings see Programming p 25 3 8 Flow Meter Connect the flow meter if any to the AMU transmitter according to the connection diagram see Electrical Connection p 19 3 9 Input NO...

Страница 24: ...1 Signal Output 1 and 2 current outputs NOTICE Max burden 510 Ω If signals are sent to two different receivers use signal isolator loop isolator Signal output 1 Terminals 13 and 12 Signal output 2 Terminals 14 and 12 For programming see Program List and Explanations p 47 Menu Installation Terminals Description NC a Normally Closed a As defined when ordering 5 6 Active opened during normal operatio...

Страница 25: ...er The RS232 interface is used for logger download and firmware up load 3 12 2 Profibus optional To connect several instruments by means of a network or to config ure a PROFIBUS DP connection consult the PROFIBUS manual Use appropriate network cable NOTICE The switch must be ON if only one instrument is installed or on the last instrument in the bus ...

Страница 26: ...llation 3 12 3 Modbus optional To connect several instruments by means of a network consult the MODBUS manual Use appropriate network cable NOTICE The switch must be ON if only one instrument is installed or on the last instrument in the bus ...

Страница 27: ...arameters see 5 1 Sensors p 50 Flow measurement Measuring Mode USP Operating Mode Sensor parameters Temperature compensation Program all parameters for external devices interface re corders etc Program all parameters for instrument opera tion limits alarms See Program Overview p 42 for explanations see Program List and Explanations p 47 Flow Measurement Menu 5 1 1 Select the flow sensor you are us...

Страница 28: ...g values are compared with the values of an implemented table defined by USP If the deviation is too high the Error 15 USP Error is issued Further information see USP Operating Mode p 8 Sensor Parameters Menu 5 1 4 Enter the following parameters printed on the sensor label Cell constant ZK Temperature correction DT Sensor cable length If the sensor cable length is 0 3 m set the cable length to 0 m...

Страница 29: ...es to move back to the previous menu level B to move DOWN in a menu list and to decrease digits C to move UP in a menu list and to increase digits D to open a selected sub menu to accept an entry Exit Enter A B C D 25 4 C RUN 80 l h 14 10 45 R1 17 04 M R2 1 Installation Operation Diagnostics Messages Maintenance Main Menu Enter Exit ...

Страница 30: ...OR Error Fatal Error C Keys locked transmitter control via Profibus D Time E Process values F Sample temperature G Sample flow H Relay status upper lower limit not yet reached upper lower limit reached control upw downw no action control upw downw active dark bar indicates control intensity motor valve closed motor valve open dark bar indicates approx position timer timer timing active hand rotati...

Страница 31: ...s that might need to be modified during daily routine Normally password protected and used by the process operator Subset of menu 5 Installation but process related Menu Installation 5 For initial instrument set up by SWAN authorized person to set all instrument parameters Can be protected by means of password 1 Messages Operation Maintenance Diagnostics Main Menu Installation 1 1 Pending Errors M...

Страница 32: ...w None Meas Mode Resistivity Sensor parameters USP Operating Mode off 5 1 1 Sensors Temp Compensation Flow None Meas Mode Resistivity Sensor parameters USP Operating Mode off Flow None Q Hflow Q flow 5 1 1 Sensors Temp Compensation Flow Q Hflow Meas Mode Resistivity Sensor parameters USP Operating Mode off 5 1 1 Sensors Temp Compensation Meas Mode Resistivity Sensor parameters USP Operating Mode o...

Страница 33: ... WARNING Electrical shock hazard Do not carry out maintenance work during normal operation Always turn off power before manipulating electric parts 1 Shut off power of the instrument 2 Stop sample flow by closing the flow regulating valve If necessary Clean sensor According to USP regulations Perform a transmitter check ...

Страница 34: ...th fixed cable To avoid damage of the sensor cable due to torsion when screwing the sensor out of the flow cell disconnect the cable from the terminals of the AMU transmitter Remove the sensor with fixed cable 1 Open the transmitter housing 2 Disconnect the sensor cable from the terminals 3 Remove the sensor cable from the transmitter housing 4 Unscrew and remove the sensor A from the flow cell bl...

Страница 35: ...rew the sensor into the flow cell and tighten it well 3 Feed the sensor cable into the transmitter housing 4 Connect the sensor cable to the terminals of the AMU transmitter see Electrical Connection S 19 5 Close the transmitter housing 6 Open the flow regulating valve C 7 Switch on power A B C Sensor Flow cell Flow regulating valve A B C ...

Страница 36: ... B 2 Proceed according to Remove the Sensor with fixed cable step 4 Install the sen sor with plug 1 Wrap 7 turns of teflon tape around the sensor thread 2 Screw the sensor into the flow cell and tighten it well 3 Screw the sensor plug on to the senor 4 Open the flow regulating valve C 5 Switch on power A B Sensor plug Sensor A B ...

Страница 37: ... 1 for conductivity NOTICE Keep test kit absolutely dry Several variables enter into the calculation of specific conductivity Values like the cell constant temperature compensation algorithm cable length and temperature correction of the sensor are stored in the memory of the conductivity instrument Test mode For a quick and easy test the AMU Rescon transmitter has the test mode Transmitter check ...

Страница 38: ...on Maintenance 6 4 2 Carry out a transmitter check Connect the test resistor 1 Unscrew and remove the sensor plug A from the sensor C 2 Screw the sensor plug onto the test resistor B A B C Sensor plug Test resistor Sensor RC U A B C ...

Страница 39: ...in the menu Maintenance Fine adjust 1 Navigate to menu Maintenance Transmitter check 2 Press Enter 3 Select on and confirm with En ter The Transmitter check is now active 4 Abandon the menu by pressing the exit key twice 5 Compare the displayed resistivity or conductivity value and the tem perature value with the value print ed on the test resistor 3 3 Set Time07 05 12 11 06 17 Maintenance Fine ad...

Страница 40: ...38 A 96 250 351 230320 AMU Rescon Maintenance 6 6 Longer Stop of Operation 1 Stop sample flow 2 Shut off power of the instrument 3 Unscrew and remove the sensor 4 Empty and dry the flow cell ...

Страница 41: ... e Sample Flow low Such Errors are marked E0xx bold and orange Errors which indicate a hardware failure of the instrument Such Errors are marked E0xx bold and red Error or fatal Error Error not yet acknowledged Check Pending Errors 1 1 5 and take corrective action Press ENTER Navigate to menu Pending Errors Press ENTER Press ENTER to acknowledge the Pending Errors The Error is reset and saved in t...

Страница 42: ...heck programmed value E011 Temp shorted Check wiring of sensor Check sensor E012 Temp disconnected Check wiring of sensor Check sensor E013 Case Temp high check case environment temperature check programmed value E014 Case Temp low check case environment temperature check programmed value E015 USP Error check process E017 Control Timeout check control device or programming in Installation Relay co...

Страница 43: ...IC LM75 call service E030 EEProm Frontend call service E031 Calibration Recout call service E032 Wrong Frontend call service E033 Power on none normal status E034 Power down none normal status E065 Transmitter check Error Description Corrective action ...

Страница 44: ...gs can be modified Menu 3 Maintenance is for service Calibration simulation of outputs and set time date Please protect with password Menu 4 Operation is for the user allowing to set limits alarm values etc The presetting is done in the menu Installation only for the System engineer Please protect with password Menu 5 Installation Defining assignment of all inputs and outputs measuring parameters ...

Страница 45: ...Front End Operating Time Years Days Hours Minutes Seconds 2 1 4 1 2 1 4 Sensors Cond Sensor Current Value MOhm 2 2 2 2 1 Raw value MOhm Cell Constant Cal History Number 2 2 1 5 1 2 2 1 5 Date Time RSlo KOhm Miscellaneous Case Temp 2 2 2 1 2 2 2 Sample Sample ID 2 3 1 2 3 Temperature C Nt5K Ohm I O State Alarm Relay 2 4 1 2 4 Relay 1 2 2 4 2 Input Signal Output 1 2 Interface Protocol 2 5 1 2 5 Baud...

Страница 46: ... adjust Current Value 3 5 1 3 5 RSlo Sensors Filter Time Const 4 1 1 Menu numbers 4 1 Hold after Cal 4 1 2 Relay Contacts Alarm Relay Alarm Alarm High 4 2 1 1 1 4 2 4 2 1 4 2 1 1 Alarm Low 4 2 1 1 26 Hysteresis 4 2 1 1 36 Delay 4 2 1 1 46 Relay 1 and 2 Setpoint 4 2 x 100 4 2 2 and 4 2 3 Hysteresis 4 2 x 200 Delay 4 2 x 30 Input Active 4 2 4 1 4 2 4 Signal Outputs 4 2 4 2 Output Control 4 2 4 3 Fau...

Страница 47: ...tion 5 2 1 3 5 2 2 3 Scaling Range Low 5 2 x 40 10 10 5 2 x 40 Range High 5 2 x 40 20 20 Relay Contacts Alarm Relay Alarm Alarm High 5 3 1 1 1 5 3 5 3 1 5 3 1 1 Alarm Low 5 3 1 1 26 Hysteresis 5 3 1 1 36 Delay 5 3 1 1 46 Sample Flow Flow Alarm 5 3 1 2 1 5 3 1 2 Alarm High 5 3 1 2 2 Alarm Low 5 3 1 2 36 Sample Temp Alarm High 5 3 1 3 1 5 3 1 3 Alarm Low 5 3 1 3 26 Case Temp high 5 3 1 4 Case Temp l...

Страница 48: ...2 Output Control 5 3 4 3 Fault 5 3 4 4 Delay 5 3 4 5 Miscellaneous Language 5 4 1 5 4 Set defaults 5 4 2 Load Firmware 5 4 3 Password Messages 5 4 4 1 5 4 4 Maintenance 5 4 4 2 Operation 5 4 4 3 Installation 5 4 4 4 Sample ID 5 4 5 Interface Protocol 5 5 1 5 5 Device Address 5 5 21 Baud Rate 5 5 31 Parity 5 5 41 ...

Страница 49: ...r 2 Diagnostics In diagnostics mode the values can only be viewed not modified 2 1 Identification Desig Designation of the instrument Version Firmware of the instrument e g V6 20 11 16 2 1 3 Factory Test Test date of the instrument and motherboard 2 1 4 Operating Time Shows the operating time in years days hours minutes and seconds 2 2 Sensors 2 2 1 Cond Sensor 2 2 1 5 Cal History Only visible in ...

Страница 50: ...in 2 4 I O State Shows actual status of all in and outputs 2 4 1 2 5 Interface Only available if optional interface is installed Review programmed communication settings 3 Maintenance 3 1 Simulation To simulate a value or a relay state select the alarm relay relay 1or 2 signal output 1or 2 with the or key Press the Enter key Change the value or state of the selected item with the or key Press the ...

Страница 51: ... the AMU Rescon is set to the measuring mode resistivity The function fine adjust performs an internal resistor adjustment 4 Operation 4 1 Sensors 4 1 1 Filter Time Constant Used to damp noisy signals The higher the filter time constant the slower the system reacts to changes of the measured value Range 5 300 Sec 4 1 2 Hold after Cal Delay permitting the instrument to stabilize again af ter calibr...

Страница 52: ... logging buffer is full the oldest data record is erased to make room for the newest one circular buffer Range 1 Second to 1 hour 4 3 2 Clear Logger If confirmed with yes the complete logger data is de leted A new data series is started 5 Installation 5 1 Sensors 5 1 1 Flow Select the type of flow cell Possible flow cells Selection of the suitable flow cell 5 1 2 Meas Mode The two measuring modes ...

Страница 53: ...rity water strong acids strong bases ammonia ethanolamine morpholine 5 2 Signal Outputs NOTICE The navigation in the menu Signal Output 1 and Signal Output 2 is identical For reason of simplicity only the menu numbers of Signal Output 1 are used in the following 5 2 1 and 5 2 2 Signal Output 1 and 2 Assign process value the current loop range and a function to each signal output 5 2 1 1 Parameter ...

Страница 54: ...ntrollers See As control output p 53 As process values The process value can be represented in 3 ways linear bilinear or logarithmic See graphs below 5 2 1 40 Scaling Enter beginning and end point Range low high of the linear or logarithmic scale In addition the midpoint for the bilinear scale If Parameter Meas Value A B linear bilinear X Measured value X Measured value logarithmic 20 0 0 0 1 0 2 ...

Страница 55: ... is characterized by the P Band In the steady state the setpoint will never be reached The deviation is called steady state error Parameters setpoint P Band PI controller The combination of a P controller with an I controller will minimize the steady state error If the reset time is set to zero the I controller is switched off Parameters setpoint P Band reset time PD controller The combination of ...

Страница 56: ...is reduced from 100 to 0 to reach the set point without overshooting 5 2 1 43 Control Parameters if Parameter Meas value 5 2 1 43 10 Setpoint 0 00 200 M or 0 000 2000 S 5 2 1 43 20 P Band 0 00 200 M or 0 000 2000 S 5 2 1 43 Control Parameters if Parameter Temperature 5 2 1 43 11 Setpoint 30 to 130 C 5 2 1 43 21 P Band 0 to 100 C 5 2 1 43 Control Parameters if Parameter Sample Flow 5 2 1 43 12 Setp...

Страница 57: ...afety reasons Range 0 720 min 5 3 Relay Contacts 5 3 1 Alarm Relay The alarm relay is used as cumulative error indicator Under normal operating conditions the contact is active The contact is inactive at Power loss Detection of system faults like defective sensors or electronic parts High case temperature Process values out of programmed ranges Program alarm levels for the following parameters Mea...

Страница 58: ...flow is essential for a correct measurement We recommend to program yes 5 3 1 2 2 Alarm High If the measuring values rises above the programmed value E009 will be issued Range 0 200 l h 5 3 1 2 36 Alarm Low If the measuring values falls below the programmed value E010 will be issued Range 0 200 l h 5 3 1 3 Sample Temp Define at which sample temperature an alarm should be issued 5 3 1 3 1 Alarm Hig...

Страница 59: ...wer limit switches program the following 5 3 2 20 Parameter select a process value 5 3 2 300 Setpoint If the measured value rises above respectively falls below the set point the relay is activated 5 3 2 400 Hysteresis within the hysteresis range the relay does not switch This prevents damage of relay contacts when the measured value fluctuates around the alarm value 5 3 2 1 Function Limit upper l...

Страница 60: ...vices that are driven time proportional are solenoid valves peristaltic pumps Dosing is controlled by the operating time 5 3 2 32 20 Cycle time duration of one control cycle on off change Range 0 600 sec 5 3 2 32 30 Response time Minimal time the metering device needs to react Range 0 240 sec 5 3 2 32 4 Control Parameters Range for each Parameter same as 5 2 1 43 p 54 Examples of metering devices ...

Страница 61: ...Mode Operating mode 5 3 2 340 Interval The interval can be programmed within a range of 1 1 440 min 5 3 2 44 Run Time Enter the time the relay stays active Range 5 32 400 sec 5 3 2 54 Delay during run time plus the delay time the signal and control outputs are held in the operating mode programmed below Range 0 6 000 sec 5 3 2 6 Signal Outputs Select operating mode of the signal output Actuator Mo...

Страница 62: ...5 3 2 6 Signal Outputs see Interval 5 3 2 7 Output Control see Interval The relay contact can be activated at one or several days of a week The daily starting time is valid for all days 5 3 2 342 Calendar 5 3 2 342 1 Start time The programmed start time is valid for each of the pro grammed days To set the start time see 5 3 2 341 p 60 Range 00 00 00 23 59 59 5 3 2 342 2 Monday Possible settings on...

Страница 63: ...eration Range 0 6 000 Sec 5 3 2 1 Function Fieldbus No Input is never active When closed Input is active if the input relay is closed When open Input is active if the input relay is open Continuous Signal outputs continue to issue the measured value Hold Signal outputs issue the last valid measured value Measurement is interrupted Errors except fatal errors are not issued Off Set to 0 or 4 mA resp...

Страница 64: ...alues including communication parameters 5 4 3 Load Firmware Firmware updates should be done by instructed service personnel only 5 4 4 Password Select a password different from 0000 to prevent unau thorized access to the following menus 5 4 4 1 Messages 5 4 4 2 Maintenance 5 4 4 3 Operation 5 4 4 4 Installation Each menu may be protected by a different password If you forgot the passwords contact...

Страница 65: ...ters must be defined 5 5 1 Protocol Profibus 5 5 20 Device address Range 0 126 5 5 30 ID Nr Range Analyzer Manufacturer Multivariable 5 5 40 Local operation Range Enabled Disabled 5 5 1 Protocol Modbus RTU 5 5 21 Device address Range 0 126 5 5 31 Baud Rate Range 1200 115200 Baud 5 5 41 Parity Range none even odd 5 5 1 Protocol Hyperterminal Baud Rate Range 1200 115200 Baud ...

Страница 66: ...d after Cal 300 s Alarm Relay same as in Installation Relay 1 and 2 same as in Installation Input same as in Installation Logger Logger Interval 30 min Clear Logger no Installation Sensors Flow None Meas Mode remains as set Resistivity or Conductivity USP Operating Mode Off Sensor parameters Cell Constant 0 01000 cm 1 Temp Corr 0 00 C Cable length 0 0 m Temp Compensation Comp None Signal Output 1 ...

Страница 67: ...on limit upper Parameter Meas Value Resistivity Setpoint 200 MΩ Hysteresis 1 MΩ Conductivity Setpoint 1000 µS Hysteresis 10 00 µS Delay 30 s If Function Control upw or dnw Parameter Meas Value Settings Actuator Frequency Settings Pulse Frequency 120 min Resistivity Settings Control Parameters Setpoint 200 MΩ Settings Control Parameters P band 1 MΩ Conductivity Settings Control Parameters Setpoint ...

Страница 68: ...roportional Cycle time 60 s Response time 10 s Settings Actuator Motor valve Run time 60 s Neutral zone 5 If Function Timer Mode Interval Interval 1 min Mode daily Start time 00 00 00 Mode weekly Calendar Start time 00 00 00 Calendar Monday to Sunday Off Run time 10 s Delay 5 s Signal output cont Output Control cont Input Active when closed Signal Outputs hold Output Control off Fault no Delay 10 ...

Страница 69: ...cell B Flow 130 12 QV Flow QV HFlow 11 Fluidics 8 I Input 7 21 Interface 7 Interfaces Modbus 24 Profibus 23 RS232 23 L Longer Stop of Operation 38 M Modbus 24 P Power supply 21 Profibus 23 R Relays 7 RS232 23 S Sample flow establish 25 Setup 25 Signal 7 Signal Outputs 7 Software 29 Specifications AMU Transmitter 11 Swansensor RC U 13 System Description of 7 T Terminals 19 22 U USP Operating Mode 8...

Страница 70: ...68 A 96 250 351 230320 AMU Rescon Notes 12 Notes ...

Страница 71: ...AMU Rescon Notes A 96 250 351 230320 69 ...

Страница 72: ... companies and distributors cooperates with independent representatives all over the world SWAN Products Analytical Instruments for High Purity Water Feedwater Steam and Condensate Potable Water Pool and Sanitary Water Cooling Water Waste Water and Effluents Made in Switzerland ...

Отзывы: