AMU Rescon
Maintenance
6.3. Maintenance of Sensor
6.3.1
Clean Sensor
The Swansensor RC-U is largely maintenance free. However, de
-
pending on the application, it can be contaminated, which may
cause problems.
The Swansensor RC-U is available in the following 2 different ver
-
sions:
Sensor with fixed cable
Sensor with plug
NOTICE:
Sensor with fixed cable
• To avoid damage of the sensor cable due to torsion when
screwing the sensor out of the flow cell, disconnect the cable
from the terminals of the AMU transmitter.
Remove the
sensor with
fixed cable
1
Open the transmitter housing.
2
Disconnect the sensor cable from the terminals.
3
Remove the sensor cable from the transmitter housing.
4
Unscrew and remove the sensor [A] from the flow cell block [B],
use a monkey spanner.
5
Remove the teflon tape from the sensor thread.
6
Clean the sensor with soap water.
7
Rinse the sensor well with high purity water.