AMU pH-Redox
Instrument Setup
A-96.250.441 / 060619
39
Set Buffer
Values
Please note that this list is only valid for Swan buffers. If you use
different buffers please ask the manufacturer.
The temperature curves for the buffer solutions for:
Standard 1 = pH7
Standard 2 = pH9
are already implemented in the transmitter firmware. To program
the temperature curve for the buffer solution pH4 overwrite stan-
dard 2.
Temperature
Value pH7
Value pH9
Value pH4
Buffer value at 0°C
7.13
9.24
Buffer value at 5°C
7.07
9.16
3.99
Buffer value at 10°C
7.05
9.11
3.99
Buffer value at 15°C
7.02
9.05
3.99
Buffer value at 20°C
7.00
9.00
3.99
Buffer value at 25°C
6.98
8.95
4.01
Buffer value at 30°C
6.97
8.91
4.01
Buffer value at 35°C
6.96
8.88
Buffer value at 40°C
6.95
8.85
4.03
Buffer value at 50°C
6.95
8.79
4.05
Buffer value at 60°C
4.09