66
A-96.250.441 / 060619
AMU pH-Redox
Program List and Explanations
5.1.3.3
Temp. Compensation:
(only available for pH measurement)
5.1.3.3.1
Comp.: Choose the compensation model which fits best to your ap-
plication. Available compensation models:
5.1.40
Standards:
(only visible if pH is selected)
A temperature curve is programmed for SWAN standard 1, pH 7
and SWAN standard 2, pH 9. If you want to use your own standards
you can readjust the temperature curve according to your stan-
dards.
5.1.40.1
Standard 1
: Assign the measured pH value to the according tem-
perature from 0–50 °C in steps of 5 °C.
5.1.40.2
Standard 2
: Assign the measured pH value to the according tem-
perature from 0–50 °C in steps of 5 °C.
5.1.4
Standard
: (only visible if redox is selected)
Enter the mV value of the redox standard.
Comp.
We recommend
Nernst
for potable water, waste water, swimming pools
non-linear
for high purity water
coefficient
for high purity water