AMU pH-Redox
Instrument Setup
4.
Instrument Setup
4.1.
Establish Sample Flow
1
Open sample flow tap.
2
Wait until the flow cell is completely filled.
3
Switch on power.
4.2.
Programming
Programming
Set all necessary sensor parameters in menu 5.1 Installation/Sen-
sors, further information see
Type of sensor: set the type of sensor to pH or Redox accord-
ing to your application
Flow measurement: Set the flow measurement according to
the installed flow sensor.
Temperature:
If a temperature sensor is installed, set Temp. sensor to yes.
If no temperature sensor is used, set the default temperature
to the expected sample temperature.
Standard solution(s): Program the buffer values (pH buffer ta-
ble) or the ORP calibration solution if you do not use the
SWAN standard
Program all parameters for external devices (interface, recorders,
etc.). Program all parameters for instrument operation (limits,
alarms). See
Program List and Explanations, p. 61
for explana-
tions.
Calibration of
pH electrode
The instrument should be operating for 1 h before performing a pH
calibration.
Calibrate the pH electrode with two buffers, e.g. pH 7.00 and pH
, for details.
Calibration of
ORP electrode
The instrument should be operating for 1 h before performing an
, for details.