AMU pH-Redox
Program List and Explanations
A-96.250.441 / 060619
79
5.4.4
Password:
Select a password different from 0000 to prevent unau-
thorized access to the following menus:
5.4.4.1
Messages
5.4.4.2
Maintenance
5.4.4.3
Operation
5.4.4.4
Installation.
Each menu may be protected by a
different
password.
If you forgot the passwords, contact the closest SWAN representative.
5.4.5
Sample ID:
Identify the process value with any meaning full text,
such as KKS number.
5.5 Interface
Select one of the following communication protocols. Depending on
your selection, different parameters must be defined.
5.5.1
Protocol:
Profibus
5.5.20
Device address:
Range: 0–126
5.5.30
ID-Nr.:
Range: Analyzer; Manufacturer; Multivariable
5.5.40
Local operation:
Range: Enabled, Disabled
5.5.1
Protocol:
Modbus RTU
5.5.21
Device address:
Range: 0–126
5.5.31
Baud Rate:
Range: 1200–115 200 Baud
5.5.41
Parity:
Range: none, even, odd
5.5.1
Protocol:
Hyperterminal
Baud Rate:
Range: 1200–115 200 Baud