26
A-96.250.441 / 060619
AMU pH-Redox
Installation
Install the
sensor
This instruction applies for both the pH and the ORP sensor.
1
Unscrew and remove the blind plug [F] from the flow cell block.
2
Carefully remove the protective cap [I] from the sensor tip. Turn
it clockwise only.
3
Rinse the sensor tip with clean water.
4
Insert the sensor through the flow cell block [H] into the calibra-
tion vessel [J].
5
Tighten it hand-tight.
6
Remove the connector cap [B].
7
Screw the connector [E] onto the sensor.
8
Keep the protective caps on a secure place for later use.
9
Attach the KCl supply pipe to the dosing tip of the KCl bottle.
10
Mount the KCl bottle to the bottle holder.
A
B
C
D
E
KCl bottle
Connector cap
KCl supply pipe
Dosing tip
Connector
F
G
H
I
J
Blind plug
Sensor
Flow cell block QV-Flow
Protective cap
Calibration vessel
B
A
F
G
H
J
I
E
C
D