AMU pH-Redox
Maintenance
Clean
pH Electrode
1
If necessary wipe the electrode shaft and the green tip cautious-
ly with a soft, clean, and damp paper tissue.
2
Remove grease with a tissue moistened with alcohol.
3
Slightly open the sensor cap of the reference sensor and allow a
small amount of electrolyte (~5 ml) to flow out.
4
Tighten the sensor cap hand tight again.
5
Rinse the electrode tip thoroughly with clean water.
6
Install the electrode into the flow cell again.
7
Let the electrode run-in for 1 h before the first calibration.