Program List and Explanations
54
3.30 Simulation
To simulate a value or a relay state, select the
alarm relay,
relay 1 and 2
signal output 1 and 2
with the [
] or [
] key.
Press the <Enter> key.
Change the value or state of the selected item with the [
] or
[
] key.
Press the <Enter> key.
The value is simulated by the relay/signal output.
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val
-
ues will be reset.
3.4 Replace Absorber
If maintenance message E068 or error message E006 appears, call
this menu item and replace the humidity absorber. This will make
both messages disappear for 3 days. This time is needed for the new
humidity absorber to dry the housing.
3.5 Set Time
Adjust date and time.
Alarm Relay:
Active or inactive
Relay 1 and 2:
Active or inactive
Input:
Open or closed
Signal Output 1 and 2:
Actual current in mA
Signal Output 3:
Actual current in mA (if option is installed)