2.1. Instrument Specification
CAUTION
The measuring chamber is under the pressure of the
sample line
If the inlet pressure is too high, moisture can enter the housing of
the measuring electronics and lead to incorrect measurements
and failure of the instrument.
The maximum inlet pressure of 10 bar must never be
exceeded, not even for short time periods. Take appropriate
measures to protect the instrument from pressure surges in the
sample line.
Power Supply
AC variant:
DC variant
Power consumption:
100–240 VAC (± 10%)
50/60 Hz (± 5%)
10–36 VDC
max. 35 VA
Transmitter
specifications
Housing:
Ambient temperature:
Storage and transport:
Humidity:
Display:
aluminum, with a protection degree of
IP 66 / NEMA 4X
−10 to +50 °C
−30 to +85 °C
10–90% rel., non condensing
backlit LCD, 75 x 45 mm
Sample
requirements
Flow rate:
Temperature:
Inlet pressure
Outlet pressure:
5–20 l/h
1–40 °C
1–10 bar
pressure free
On-site
The analyzer site must permit connections to:
requirements
Sample inlet:
Sample outlet:
Serto 4x6 mm
Hose nozzle, 15x20 mm