To start a verification proceed as follows:
If the value is lower or higher than ±10%, proceed as follows:
Check the sample temperature and if necessary adjust it.
Clean the measuring chamber.
1
Navigate to menu <Maintenance>/
<Verification> and press [Enter].
2
Stop sample flow.
3
Empty measuring chamber by
turning the lever on the solenoid
valve upwards.
4
Open the measuring chamber.
5
Insert the test unit with the middle
flat area pointing towards to you
into the measuring chamber of the
AMI Turbitrack.
6
Open the sample inlet valve.
7
Press [Enter].
Setpoint is displayed.
8
Press [Enter].
9
Enter the setpoint printed on the la
-
bel of the test unit with the [
]
or [
] key.
10
Press [Enter] to confirm.
The verification is now running.
If the deviation is within ±10%, press
[Enter] to save the value.
3.1
Rinsing
Maintenance
Set Time 01.01.05 16:30:00
Simulation
Verification
3.1.5
- Close inlet valve
- Open inlet valve
Verification
- Replace Measuring Unit
with Test Unit
<Enter> to continue
3.1.5
Verification
Setpoint
5.00 FNU
<Enter> to continue
3.1.5
- Actual Value
x.xx FNU
Verification
- Setpoint
5.00 FNU
- Deviation
x.xx%
<Enter> to save