AMI Deltacon DG
Troubleshooting
46
E012
Temp. 1 disconnected
– Check wiring of temperature sensor
– Check temperature sensor
E013
Case Temp. high
– check case/environment temperature
– check programmed value, see
E014
Case Temp. low
– check case/environment temperature
– check programmed value, see
E015
pH Calculation undef.
– pH out of range (<7.5 or >11.5)
E017
Control time-out
– Check control device or programming
in Installation, Relay contact, Relay 1/2
E018
Degasser disconnected
– check wiring of degasser
E019
Temp. 2 shortened
– check wiring of temperature sensor
– check temp. sensor
E020
Temp. 2 disconnected
– check wiring of temperature sensor
– check temp. sensor
E021
Temp. 3 shortened
– check wiring of temperature sensor
– check temp. sensor
E022
Temp. 3 disconnected
– check wiring of temperature sensor
– check temp. sensor
E023
Degasser Ctl Timeout
–
E024
Input active
– See If Fault Yes is programmed in
E026
IC LM75
– call service
E028
Signal output open
– check wiring on signal outputs 1 and 2
E029
Calibration degasser
– call service
E030
EEProm Frontend
– call service
E031
Cal. Recout
– call service
E032
Wrong Frontend
– call service