AMI Deltacon DG
Program List and Explanations
72
5.3.1.4.2
Alarm low:
Set the alarm low value for temperature of electronics
housing. If the value falls below the programmed value E014 is is
-
sued.
Range: -10 to +20 °C
5.3.2 & 3
Relay 1 and 2:
The contacts can be set as normally open or normal
-
. The function of re
-
lay contacts 1 or 2 is defined by the user.
Note:
The navigation in the menu <Relay 1> and <Relay 2> is
equal. For reason of simplicity only the menu numbers of Relay 1
are used in the following.
1
First select the functions as:
-
Limit upper/lower,
-
Control upwards/downwards,
- Timer
- Fieldbus
2
Then enter the necessary data depending on the selected func
-
tion. The same values may also be entered in menu
When the relays are used as upper or lower limit switches, program
the following:
5.3.2.20
Parameter:
select a process value
5.3.2.300
Setpoint
: If the measured value rises above respectively falls below
the set-point, the relay is activated.
5.3.2.1
Function = Limit upper/lower:
Parameter
Range
Cond. 1 (sc)
0.000–3000
S
Cond.2 (cc)
0.000–3000
S
Cond.3 (dc)
0.000–3000
S
Temp. 1
-25 to +270 °C
Temp. 2
-25 to +270 °C
Temp. 3
-25 to +270 °C
Difference
0.000–3000
S
Sample flow
0–20 l/h
pH
0 –14 pH
Ammonia
0 –500 ppm