AMI Deltacon DG
Instrument Setup
32
Calculations
Menu 5.1.1.1
Set <Calculations> to “Yes” if you want to have pH and alkalization
agent calculated and displayed.
Measuring unit
Menu 5.1.1.2
Set the <Measuring unit> according to your requirements:
S/cm
S/m
Monitoring of
cation ex
-
change resin
Menu 5.1.1.3
Set <Monitoring of resin> to “Yes” if you want to monitor the capacity
of the cation exchanger resin.
Display
Menu 4.4.1, Screen 1
Menu 4.4.2, Screen 2
Program display screens according to your requirements, see pro
-
gram list and explanations
External
devices
Program all parameters for external devices (interface, recorders,
etc.) See program list and explanations
Limits Alarms
Program all parameters for instrument operation (limits, alarms). See
.
Degasser
CAUTION
Possible damage of the degasser
If the instrument is completely empty when first started up, the
degasser heats up at full power for a short time without sample.
This can damage the degasser.
Open the flow regulating valve and wait until the sample flows
into the waste funnel before switching on the degasser.
To switch the degasser on navigate to menu 5.1.2.4, <Installation/
Sensors/Sensor parameters/Degasser/Mode> and set Mode to
“on”.