A-96.250.491 / 190121
AMI Deltacon DG
Program List and Explanations
59
3.3 Change of Resin
Yes or No
: After exchanging the resin bottle navigate to this menu,
press [Enter], select <Yes>, press [Exit] and confirm with <Yes>.
With this action the resin capacity is set to 100% and the current
date is saved. Date and resin capacity are displayed in the menu
<Diagnostics/Sample>
3.5 Degasser
3.5.1
Define Boiling point
In this menu you can start a manual determination of the boiling point
at any time.
4 Operation
4.1 Sensors
4.1.1
Filter Time Constant:
Used to damp noisy signals. The higher the fil
-
ter time constant, the slower the system reacts to changes of the
measured value.
Range: 5–300 Sec
4.1.2
Hold after Cal.:
Delay permitting the instrument to stabilize again af
-
ter calibration. During calibration plus hold-time, the signal outputs
are frozen (held on last valid value), alarm values, limits are not ac
-
tive.
Range: 0–6‘000 Sec
4.2 Relay Contacts
See
4.3 Logger
The instrument is equipped with an internal logger. The logger data
can be copied to a PC with an USB stick if option USB interface is in
-
stalled.
The logger can save approx. 1500 data records. The Records con
-
sists of: Date, time, alarms, measured values, measured value un
-
compensated.
Range: 1 Second to 1 hour
4.3.1
Log Interval:
Select a convenient log interval. Consult the table be
-
low to estimate the max logging time. When the logging buffer is full,
the oldest data record is erased to make room for the newest one
(circular buffer).