
A-96.250.491 / 190121
AMI Deltacon DG
Troubleshooting
44
7.1. Error List
Error
Non-fatal Error. Indicates an alarm if a programmed value is exceed
-
ed.
Such Errors are marked
E0xx
(bold and black).
Fatal Error
(blinking symbol)
Control of dosing devices is interrupted.
The indicated measured values are possibly incorrect.
Fatal Errors are divided in the following two categories:
Errors which disappear if correct measuring conditions are re
-
covered (i.e. Sample Flow low).
Such Errors are marked
E0xx
(
bold and orange
)
Errors which indicate a hardware failure of the instrument.
Such Errors are marked
E0xx
(
bold and red
)
Error or
fatal Error
Error not yet acknowledged.
Check
Pending Errors 1.1.5
and take
corrective action.
Navigate to menu <Messages>/
<Pending Errors>.
Press [ENTER] to acknowledge the
Pending Error.
The Error is reset and saved in
the Message List.
25.4°C
HOLD
8 l/h
14:10:45
R1
8.458
S
0.171
S
R2
1.1
Message List
Pending Errors
Maintenance List
Messages
1.1.5
Pending Errors
Error Code
E010
Sample Flow low
<Enter> to Acknowledge