AMI Deltacon DG
Program List and Explanations
71
5.3.1.2.2
Temp. 2
5.3.1.2.2.1
Alarm High:
If the measured value rises above the alarm high value,
the alarm relay is activated and E037, is displayed in the message
list.
Range: 30–200 °C
5.3.1.2.2.25
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E038 is displayed in the message
list.
Range: -10 to + 20 °C
5.3.1.2.3
Temp. 3
5.3.1.2.3.1
Alarm High:
If the measured value rises above the alarm high value,
the alarm relay is activated and E037, is displayed in the message
list.
Range: 30–200 °C
5.3.1.2.3.25
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E038 is displayed in the message
list.
Range: -10 to + 20 °C
5.3.1.3
Sample Flow:
Define at which sample flow an alarm should be is
-
sued.
5.3.1.3.1
Flow Alarm:
Program if the alarm relay should be activated if there is
a flow alarm. Choose between yes or no. The flow alarm will always
be indicated in the display, pending error list, saved in the message
list and the logger.
Available values: Yes or no
Note:
Sufficient flow is essential for a correct measurement. We
recommend to program yes.
5.3.1.3.2
Alarm High:
If the measuring values rises above the programmed
value E009 will be issued.
Range: 9–20 l/h
5.3.1.3.35
Alarm Low:
If the measuring values falls below the programmed val
-
ue E010 will be issued.
Range: 5–8 l/h
5.3.1.4
Case Temp.
5.3.1.4.1
Alarm high:
Set the alarm high value for temperature of electronics
housing. If the value rises above the programmed value E013 is is
-
sued.
Range: 30–75 °C