The programmed measuring intervals of tc2 [F] and fac [G] may
overlap each other.
The tc2 measurement has priority and therefore always takes place
in the programed measuring interval [F]. The fac measurement is
filled in the gaps [H] of the overall measurement and may be shifted
(orange arrow) in respect to programmed interval. If 2 fac measure-
ments are programmed during a overall measurement one of them
is cancelled (red arrow).
1
2
3
5
6
7
8
min
4
tc 2
fac
A
B
D
E
C
tc 1
G
G
J
J
fac
A
B
C
D
E
G
J
fac
Reaction time for measuring tc 2
Flush time
Overall measurement time
Flush time end
fac measurement
Valve 1 for adding OXYCON ON-LINE KI
Valve 2 for closing photometer inlet
Free available chlorine
tc1
tc2
Total chlorine 1
Total chlorine 2
OXYCON ON-LINE DPD
OXYCON ON-LINE Buffer
OXYCON ON-LINE KI
Zero point measurement
Sample measurement
Peristaltic pump active for a short
time
F
G
H
fac
fac fac
tc2
10
0
20
30
50
60
min
40