12
A-96.250.581 / 220616
AMI Codes-II CC
Product Description
Fluidics
The sample flows through the sample inlet [R] and the inlet filter [H]
into the constant head [A]. Adjust the flow regulating valve [F] so
that always a small part of the sample flows through the overflow
tube [B] into the constant head drain [Q].
Free available
chlorine
A part of the sample flows through the photometer inlet [C] into the
mixing chamber [E], where the reagents [M] and [N] are added by
the peristaltic pump [D] and mixed with the sample. The mixed
sample flows through the photometer [K] where the free available
chlorine is measured.