108
A-96.250.581 / 220616
AMI Codes-II CC
Default Values
Parameter:
.............................................................. Calc. monochl.
Settings: Actuator: ...........................................................Frequency
Settings: Pulse Frequency: ............................................ 120/min.
Settings: Control Parameters: Setpoint: .........................5.00 ppm
Settings: Control Parameters: P-band:...........................0.10 ppm
Parameter:
...............................................................Calc. comb. cl.
Settings: Actuator: ...........................................................Frequency
Settings: Pulse Frequency: ............................................ 120/min.
Settings: Control Parameters: Setpoint: .........................5.00 ppm
Settings: Control Parameters: P-band:...........................0.10 ppm
Parameter:
........................................................Calc. dichloramine
Settings: Actuator: ...........................................................Frequency
Settings: Pulse Frequency: ............................................ 120/min.
Settings: Control Parameters: Setpoint: .........................5.00 ppm
Settings: Control Parameters: P-band:...........................0.10 ppm
Parameter:
.................................................................. Temperature
Settings: Actuator: ...........................................................Frequency
Settings: Pulse Frequency: ............................................ 120/min.
Settings: Control Parameters: Setpoint: ...............................30 °C
Settings: Control Parameters: P-band:...................................1 °C
Parameter:
..................................................................Sample Flow
Settings: Actuator: ...........................................................Frequency
Settings: Pulse Frequency: ............................................ 120/min.
Settings: Control Parameters: Setpoint: ........................... 200 B/s
Settings: Control Parameters: P-band:............................... 20 B/s
Parameter:
................................................................................... pH
Settings: Actuator: ...........................................................Frequency
Settings: Pulse Frequency: ............................................ 120/min.
Settings: Control Parameters: Setpoint: ................................ 7 pH
Settings: Control Parameters: P-band:............................... 0.1 pH
Common settings:
Settings: Control Parameters: Reset time: ...........................0 Sec
Settings: Control Parameters: Derivative Time: ...................0 Sec
Settings: Control Parameters: Control Timeout:................... 0 Min
Settings: Actuator:................................................. Time proportional
Cycle time: ............................................................................. 60 s
Response time: ..................................................................... 10 s
Settings: Actuator.......................................................... Motor valve
Run time: ............................................................................... 60 s
Neutral zone: ...........................................................................5%