AMI Codes-II CC
Program List and Explanations
A-96.250.581 / 220616
91
Control upwards/downwards
Setpoint:
User-defined process value (Measured value or flow)
P-Band:
Range below (upwards control) or above (downwards
control) the set-point, within the dosing intensity is reduced from
100% to 0% to reach the set-point without overshooting.
5.2.1.43
Control Parameters: if Parameters = Free av. chlorine
5.2.1.43.10
Setpoint
: 0–10 ppm or 0–10 mg/l
5.2.1.43.20
P-Band
: 0–10 ppm 0–10 mg/l
5.2.1.43
Control Parameters: if Parameters = Total chlorine 1
5.2.1.43.11
Setpoint
: 0–10 ppm or 0–10 mg/l
5.2.1.43.21
P-Band
: 0–10 ppm 0–10 mg/l
5.2.1.43
Control Parameters: if Parameters = Total chlorine 2
5.2.1.43.12
Setpoint
: 0–10 ppm or 0–10 mg/l
5.2.1.43.22
P-Band
: 0–10 ppm 0–10 mg/l
5.2.1.43
Control Parameters: if Parameters = Calc. monochl.
5.2.1.43.13
Setpoint
: 0–10 ppm or 0–10 mg/l
5.2.1.43.23
P-Band
: 0–10 ppm 0–10 mg/l
5.2.1.43
Control Parameters: if Parameters = Calc. comb. cl.
5.2.1.43.14
Setpoint
: 0–10 ppm or 0–10 mg/l
5.2.1.43.24
P-Band
: 0–10 ppm 0–10 mg/l
5.2.1.43
Control Parameters: if Parameters = Calc. dichloramine
5.2.1.43.15
Setpoint
: 0–10 ppm or 0–10 mg/l
5.2.1.43.25
P-Band
: 0–10 ppm 0–10 mg/l
5.2.1.43
Control Parameters: if Parameters = Temperature
5.2.1.43.16
Setpoint
: -30 to +120 °C
5.2.1.43.26
P-Band
: 0–100 °C
5.2.1.43
Control Parameters: if Parameters = Sample Flow
5.2.1.43.17
Setpoint
: 0 –600 B/s
5.2.1.43.27
P-Band
: 0 –200 B/s
5.2.1.43
Control Parameters: if Parameters = pH
5.2.1.43.18
Setpoint
: 0 –14 pH
5.2.1.43.28
P-Band
: 0 –14 pH