     SV 277 PRO User Manual 

32 

available functionalities of the station remotely accessed. 

The functionalities that a mobile modem offers are: 

  Main communication channel 

  SMS / E-mail alarms 

  SMS command exchange. 

 

4.2.1.  Main communication channel 

Main  communication  channel  is  the TCP  connection  (a  lossless data  exchange  protocol) 
that can be used to exchange commands as specified by Appendix A to SVAN 977 User 
Manual.  +  assures  this  connection  and  provides  data  download,  performance 
validation and measurement start/stop. 

Main communication channel of SV 277 PRO can be established by one of two available 
methods:  TCP  Client  or  TCP  Server.  The  SV  277 PRO  stations  support  SSL  (Secure 
Socket Layer) connections. 

The

 TCP Client 

is a mode of main communication channel in which the mobile modem is 

configured  to  initiate  connection  to  a  designated  address  (

remote  host

).  Each  time  a 

reconnection period 

(default value is 30 seconds) has elapsed, SV 277 PRO attempts to 

establish  a  TCP  connection  to  a  designated  address  on  a  designated  port  (

Data  Port

).  

Should  the  connection  be  established  successfully,  SV 277 PRO  can  exchange 
commands as in TCP Server mode. Should the connection attempt fail or is broken by the 

remote host

, SV 277 PRO will attempt to connect again after another reconnection period 

has elapsed.  

To prevent the connections from going 

idle 

(a state in which the connection seems active, 

but no data can be transferred 

– it may happen due to mobile operator terminating the link, 

but  the  TCP  protocol  not  being  able  to  detect  it),  the  station maintains  the  connection  to 
the server by sending small packages of data at reconnection period (which by default is 
one minute). If the transfer is not properly acknowledged by the other party, the connection 
will be terminated. 

 

Note: TCP Client mode is used in the SvanNET service. 

SV 277 PRO  uses  the TCP  Client mode  for connection  with  SvanNET  (this  is  the default 
setting  of  SVAN 977).  The  user  PC  connects  to  SvanNET  via  a  web  browser  or 
+, and the service creates a "bridge" between the station and the user PC. In this 
case, the user does not need to buy more options for SIM card (public dynamic or public 
static IP addresses are typically optionally payed and often hard to obtain). The essence of 
SvanNET is to simplify the procedures and requirements necessary for the connection. 

TCP  Server 

is  a  mode  in  which  the  mobile  modem  is  configured  to  act  as  a  server  for 

incoming connections. The modem is waiting for the first connection to be established on a 
designated  port  (called 

Data  Port

;  default  8001).  Such  connection  can  come  from  any 

application - an initiator of the TCP connection (such as +) called 

remote peer

This mode is recommended as it gives the initiative to the application. This mode requires 
the SIM with a 

public address 

(also called 

public IP

). 

The 

TCP Server 

mode uses a so-called 

Registration

. This is a method of translation of a 

Содержание SV 277 PRO

Страница 1: ...USER MANUAL SV 277 PRO MONITORING STATION Warsaw 2021 05 15 Rev 02 00 Copyright 2021 SVANTEK All rights reserved...

Страница 2: ...is prohibited Trademarks Trademarks or registered marks in this manual belong to their respective manufacturers Microsoft and Windows are registered trademarks of Microsoft Corporation Disclaimer Info...

Страница 3: ...he internal and external batteries as often as possible this will extend battery life It is necessary to charge the battery after any total discharge Monitoring station and or SB 272 should not be sto...

Страница 4: ...modem 16 2 1 5 Outdoor microphone protection kit 17 2 1 6 External power supply with AC DC converter 18 2 2 Optional accessories for SV 277 PRO 19 2 2 1 Sound Calibrator 19 2 2 2 External rechargeable...

Страница 5: ...mmunication module of SvanPC 34 5 svannet user interface 35 5 1 Station list view 35 5 1 1 STATUS view 37 5 1 2 LOG views 40 5 2 WEB INTERFACE view 41 5 2 1 Live data view 41 5 2 2 Configuration views...

Страница 6: ...or time constants for example Fast Slow Impulse The SA 277 outdoor microphone kit protects the SVAN 977 preamplifier and microphone from weather conditions SA 277 is made of lightweight materials and...

Страница 7: ...t The station uses a waterproof charger that is designed for the outdoor use Military standard connectors provide reliable robust and waterproof cable connections 1 2 ACCESSORIES INCLUDED SVAN 977 Cla...

Страница 8: ...option SF 277_4 FFT analysis option SF 277_15 Time domain waveform signal recording option Note The software options for the instrument can be purchased at any time as only the introduction of a spec...

Страница 9: ...nveniently into a second carrying case The SV 277 PRO station includes outdoor microphone kit 1 waterproof case with battery charging unit and connectors 2 controller 3 SVAN 977 Class 1 Sound Level Me...

Страница 10: ...ast with adjustable height from 1 5 meter to 4 meters SA 206 Type Manfrotto 269BU 2 1 1 Waterproof case The SM 277 PRO waterproof case IP 67 houses and protects the main elements of the monitoring sta...

Страница 11: ...eason Double check that the controller has a good fixation in the connector after reconnecting The monitoring station case is equipped with the air pressure compensation valve that enables the user to...

Страница 12: ...ot remove the battery from the case This operation must be done only by the authorised service Note Battery is not restricted for air surface and water transport Classified as non hazardous material I...

Страница 13: ...roller also analyses the temperature condition inside the monitoring station case If the internal temperature is higher than 50o C the controller will switch off internal battery charging If the tempe...

Страница 14: ...te of downloading the program While downloading the TEST LED starts blinking faster After successful loading the File loaded is displayed and TEST LED starts blinking red 1 per 3 seconds 4 Press any k...

Страница 15: ...e monitoring station is powered from an external power supply via the 7 16V socket Data transfer to the modem is done via the Serial socket the serial interface Other instrument sockets are not used i...

Страница 16: ...e application environmental acoustic signal is parallel to the microphone s grid or airport acoustic signal is perpendicular to the microphone s grid The frequency characteristics of the compensation...

Страница 17: ...the SVAN 977 preamplifier and microphone from weather conditions The use of the outdoor kit requires an extension cable between the instrument and its preamplifier SC 277 The SA 277 is made of lightwe...

Страница 18: ...e caps on the four antibird spikes are removed It is recommended to use the protective caps during transportation storage and other operations with the instrument like laboratory calibration etc to av...

Страница 19: ...nger operating time of the monitoring station in comparison to the station s internal 17 Ah battery The SB 272 set includes the SB 273 indoor charger and a cable for connection between SB 272 and the...

Страница 20: ...l above the monitoring case which provides a natural shade that protects the station from overheating The SB 271 is equipped with a military standard connector cable for direct connection to the monit...

Страница 21: ...econds only when it is raining or hailing These parameters are not averaged but integrated Note Do not forget to align the transmitter so that the arrow on the bottom of the transmitter points North N...

Страница 22: ...in the Logger Results list path Menu Measurement Logging Logger Results SVAN 977 also integrates weather parameters for the period defined by the parameter Integration Period in SVAN 977 path Menu Mea...

Страница 23: ...el by a combination of DC CHARGING BAT 1 or BAT 2 LEDs see Table below Note It is recommended that the batteries of SV 277 PRO and SB 272 are charged before going on site Note Monitoring station and e...

Страница 24: ...g colour is changed red orange green green when charged BAT 2 off off green if charged 50 orange if charged 20 50 red if charged 20 blinking red if charged 8 off if 0 charged or disconnected off Colou...

Страница 25: ...are powered continuously SVAN 977 instrument controller mobile modem and outdoor microphone optionally also the meteo module The power supply for these modules can be switched off only by turning off...

Страница 26: ...only if USB driver for SVANTEK instruments had been previously installed on the PC USB driver can be downloaded from www svantek com 3 4 ESTABLISHING MOBILE CONNECTION It is strongly advised to confi...

Страница 27: ...ave changed factory settings you always may restore them with the use of Factory Settings option Note If above steps don t result in connection with SvanNET or you wish to connect to the other server...

Страница 28: ...p the key on the plug and the INPUT socket then lock the connector by turning the ring clockwise only the ring close to the socket will rotate 6 If you wish to power the station connect the power supp...

Страница 29: ...urn on automatically System integration is checked directly after turning on the controller This procedure includes checking the connections with the instrument and mobile modem optionally also the me...

Страница 30: ...tek monitoring stations SvanNET allows usage of all type of SIM cards with the station mobile modem regardless of having a public or private IP The connection over the SvanNET allows users to watch re...

Страница 31: ...cess SvanNET log in to your account at https www svannet com panel login php Before logging select your language Once logged in you can use the web interface to work remotely with the monitoring stati...

Страница 32: ...state in which the connection seems active but no data can be transferred it may happen due to mobile operator terminating the link but the TCP protocol not being able to detect it the station maintai...

Страница 33: ...quire any payments for the account and obtaining any credentials as it comes preconfigured to the user but it requires access to www svantek com domain from the Mobile operator s network Note To use t...

Страница 34: ...d as specified by the said document SV277 PRO can provide a response This feature is particularly useful when retrieving the current state of the station Like SMS alarming this feature does not requir...

Страница 35: ...is manual describes only the Station list tools To get more information about Project list see SvanNET User Manual 5 1 STATION LIST VIEW Station list displays all stations assigned to your account tur...

Страница 36: ...on the appropriate button it will change its colour to blue and click it The WEB INTERFACE button switches you to the Live data view see Chapter 5 2 in which you can view measurement results and use...

Страница 37: ...LARM in the pop up box and a new Alarm 1 with CONDITIONS ACTIONS and MEASUREMENT POINTS areas will appear Alarms are based on Conditions and relate to Actions that are default e mails to the specified...

Страница 38: ...rm is generated when the system detects a decrease in the percentage of battery charge below the selected threshold Storage memory o Trigger Value xx MB GB alarm is generated when the system detects a...

Страница 39: ...et based on owner s time zone measurement is resumed Station is disconnected o Trigger value xx minutes xx hours alarm is generated when the station remains disconnected from SvanNET for a time equal...

Страница 40: ...fresh the log select the required period of records to be displayed and rewind records Data Connection log which registers history of station connections In the upper line you can refresh the log sele...

Страница 41: ...you to the Storage view see Chapter 5 2 4 in which you can download files manually Note Content of the Configuration tabs depends on the selected parameters The objective of this manual is not to pres...

Страница 42: ...uch results as Lpeak Lmax Lmin or Leq have in their names also filter abbreviation A B or Z and Lmax Lmin results have also detector type abbreviation F Fast S Slow I Impulse All results and formulas...

Страница 43: ...lues for this time point 4 Scroll the time window over the time history There are four results Leq Lpeak Lmax and Lmin measured with appropriate weighting filters and detector types and available for...

Страница 44: ...an 1 Select the Measurement function Level Meter Octave 1 1 Octave 1 3 2 Update the Instrument clock 3 Select the type of RMS LEQ Integration Linear or Exponential 4 Set the start synchronisation peri...

Страница 45: ...in a file Save summary results define measurement time and step of saving Integration period select number of measurement repetitions Repetition cycles and define ten Statistical levels 4 Configure sa...

Страница 46: ...iod during which Summary results are measured integrated with filters and time constant defined in the MEASUREMENT SETUP tab and saved in a file as the set of Summary Results The Summary step can be s...

Страница 47: ...ts are saved in the same file with Time history results Note All files with measurement result are automatically named in accordance with the rule some prefix string of letters and number string of di...

Страница 48: ...the logger file Marker audio signal recording to the wave file Audio generation of the alarm signal on the I O socket I O alarm alarm SMS sending SMS alarm or alarm e mail sending E mail alarm 3 Switc...

Страница 49: ...DS button the TIME CONDITION configuration box will pop up In this box you can select days and periods during which events occurrences will be registered After confirmation OK the pop up box closes an...

Страница 50: ...n select as a Source different results Leq Lpeak Lmax Lmin etc in the Sound mode or PEAK MAX RMS etc in the Vibration mode for three profiles Pr 1 Pr 2 and Pr 3 measured by 1s integration period for e...

Страница 51: ...lock type The Point marker means that it will be registered only at the beginning of the Event The Block marker will be registered at the beginning and at the end of the Event The Audio action starts...

Страница 52: ...rom the ADDRESS BOOK after clicking field Min break defines the minimum time break between two consequent E mails After confirmation OK the pop up window closes and the selection will be presented in...

Страница 53: ...ents a list of files saved in the instrument s SD card memory The list includes only files from a single directory on the memory card and it initially shows the content of the current working director...

Страница 54: ...ture is 20 C to 50 C 5 Storage temperature range SM 277 PRO and SA 250 40 C to 70 C 6 Working relative humidity range 0 100 RH Power Supply 7 Internal battery Europower EP 17 12 12 V 17 Ah 8 SV 277 PR...

Страница 55: ...5 0 5V and power in the range of 30W to 120 W without control system Note Size and power of the panel depend on the climate of the area where the station operates Measurement Line 16 SVAN 977 Meter a...

Страница 56: ...modem approvals R TTE GCF CE FCC PTCRB IC UL AT T and other local approvals and provider Certifications EHS6 T only 4G modem 25 modem type and features Cinterion PLS62T W Gateway LTE 1 2 3 4 5 7 8 12...

Страница 57: ...amp 2 halves 5 upper cone 6 microphone protecting tube 7 anti bird spikes 8 windscreen SA 209 9 desiccator not included 10 M5x10 hex socket raised screw 11 washer Tools K 3mm Allen key Spare parts O1...

Страница 58: ...SV 277 PRO User Manual 58 12 bottom 13 thread adapter not included O ring gasket locations inside...

Страница 59: ...SV 277 PRO User Manual 59 B 2 ASSEMBLY OF SA 277 1 Secure preamplifier A with the dummy microphone 2 and connect the cable C 2 Pull tube 1 over the cable...

Страница 60: ...lves 4 to the connector and insert O ring gasket 3 on the lightning signed place 4 Apply the upper conical part 5 on the preamplifier Note Do not use preamplifier without the dummy microphone It can c...

Страница 61: ...SV 277 PRO User Manual 61 5 Screw the tube on the cone by rotating the tube so as to not twist the cable 6 Replace the dummy microphone 2 with the microphone cartridge B...

Страница 62: ...SV 277 PRO User Manual 62 7 Calibrate the complete system using SVANTEK acoustic calibrator 8 Install the microphone protection tube 6 Use the Allen key K as a pin at protecting tube orifice...

Страница 63: ...SV 277 PRO User Manual 63 9 Mount windscreen 8 and the anti bird spikes 7 10 Slide the desiccator 9 inside the tube 11 Place the cable in the slot and connect the bottom part 12 by three screws 10...

Страница 64: ...277 PRO User Manual 64 with washers 11 Use the Allen key K 12 Mount the completed system on the dedicated tripod or pole Use thread adapter 13 if necessary see last note on next page B 3 IMPORTANT NOT...

Страница 65: ...en examining the foam as squeezing it too aggressively can damage the surface Colour of the silica gel used in the desiccator part 9 when it is dry is amber If colour of the silica gel turns light gre...

Страница 66: ...svantek com 2 SvanNET User Manual www svantek com 3 SvanPC User Manual www svantek com 4 Gemalto EHS6T Terminal User Manual www thalesgroup com IoT 5 Cinterion PLS62T W Gateway www thalesgroup com Io...

Отзывы: