![Svantek SV 277 PRO Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/svantek/sv-277-pro/sv-277-pro_user-manual_1409267029.webp)
SV 277 PRO User Manual
29
9. Take out from the case with accessories the anti-
bird spike with the foam from the accessories
case, push the foam upwards and mount the tube
on the thread. Then slide the foam down so it
hides the metal parts of the kit.
Note: Make sure that the protective caps on the four antibird spikes are
removed.
10. The station now is ready for use.
3.6. T
URNING ON THE STATION
When the remote communication is installed, and the station is fully assembled turn on the
SVAN 977 instrument. The controller will turn on automatically.
System integration is checked directly after turning on the controller. This procedure
includes checking the connections with the instrument and mobile modem (optionally also
the meteo module). Additionally, system power, internal temperature of the station and the
battery state are checked.
During the test procedure, the controller's LEDs light up sequentially orange in the
sequence from
DC
to
BAT2
and from
SVAN
to
EXT I/O
. In the case of the SD270A
controller, during the test (about 3-5s) the red LED next to the TEST button is still lit. All
errors are indicated on the front panel by appropriate LED indicators. If all red indicators
are off after above procedure, it means that the station is ready to use.