SV 277 PRO User Manual
38
2. click
OK
and new condition will be
displayed in the CONDITIONS area.
The SvanNET alarms have next meanings:
•
Mains
o
Trigger Value: Off
– alarm is generated when the system detects loss of power
supply
o
Trigger Value: On
– alarm is generated when the system detects appearance
of power supply
•
External voltage
o
Trigger Value: xx.xx V
– alarm is generated when the system detects an
external power drop below the selected value. In this case, external power
means power supply and all various battery packs
•
Battery charge
o
Trigger Value: xx % - alarm is generated when the system detects a decrease
in the percentage of battery charge below the selected threshold.
•
Storage memory
o
Trigger Value: xx MB/GB - alarm is generated when the system detects a
decrease in the free storage memory below the selected threshold.
•
System check (if applicable)
o
Alarm is generated when the system detects failure in execution of the system
check procedure (not live check).
•
Measurement stopped
o
Alarm is generated when the system detects luck of measurement. Applies
only to stopped measurements - states such as start delay, waiting for
synchronization and pause are treated as a running measurement
o
Instrument action: Start measurement
•
Storage error
o
Alarm is generated when the system detects an SD-card error. The check
1
2
a
b