SV 277 PRO User Manual
55
No.
Parameter
Value/ Description
10
SV 277 PRO power
consumption
including charging
up to 30W ±5%
11
External DC input
voltage: 10,5 V to 28 V
Note: When external DC input voltage is in the range 11 V to
15 V, the station is powered from the external DC source, but
the internal battery is not charged!
12
External DC power
supply SB 270
15 V, 4 A, 60 W
13
External
battery
SB 272
Europower EP 33-12 (12 V, 33 Ah)
14
Operating time when
powered from the
internal and external
SB272 batteries,
(20
C, 17+33 Ah,
fully charged)
Modem is switched off:
~620h (~26,0d)
Mean infrequent short transmissions: ~380h (~16,0d)
Continuous transmission:
~300h (~12,5d)
15
Solar panel (option) System best fit with solar panel with a MPPV voltage
17,5
0,5V and power in the range of 30W to 120 W (without
control system).
Note: Size and power of the panel depend on the climate of
the area where the station operates.
Measurement Line
16
SVAN 977
Meter/analyser
Datasheets are available at
17
SA 277 outdoor
microphone kit
Set of weather protection for SV 12L preamplifier and ACO
7052E microphone.
18
ACO 7052E
microphone
Pre-
polarised 1/2” condenser microphone with sensitivity 35
mV/Pa.
Note: With the use of Outdoor Airport type compensation filter (filter must be
defined in SVAN 977, see Chapter
) the acoustic characteristics of SA 277
conforms the class 1 sound meters and IEC 61672:2002 standard for the direction 0
degree (normal direction regarding the microphone membrane). With the use of
Outdoor Environmental type compensation filter the acoustic characteristics of
SA 277 is switched to the direction 90 degree (parallel to the microphone
membrane).