43
Solution: Remove the sensor from the patient. Redisplay the last measurement prior
to the timeout, or take a new measurement by placing the sensor on the
patient.
Status Code: 02
Problem: Poor sensor position (signal is inadequate for a reliable measurement)
Solution: Adjust position of sensor on patient by placing sensor on opposite hand or
ear or alternate site. Avoid fingers with nail polish or artificial nails.
Status Code: 05
Problem: The sensor has been disconnected from the device.
Solution: Reconnect the sensor. If you wish, you may leave the sensor disconnected
as this code is only displayed once at the time the sensor is disconnected.
Status Code: 10
Problem: Defective sensor
Solution: Replace the sensor and take a new measurement.
Status Code: 20
Problem: Hardware error
Solutions:
1.
Take another measurement.
2.
If this is unsuccessful, power down, then power up the unit using the power
button on the right side of the unit. If the error does not recur immediately,
take another measurement.
3.
If the error recurs, remove power from the unit (unplug any power supplies
and remove the battery, if there is one), then reconnect the power. If the
error does not recur immediately, take another measurement.
4.
If the error recurs, contact
SunTech Medical Inc.
or an authorized service
center.
Problem: Inadequate signal strength
Solution: If there is no measurement and the signal quality is low, try a different site or
sensor. Avoid fingers with nail polish or artificial nails. If there is no improvement in
signal quality, then discontinue use.